LinkSymphoKnee Distal Femoral Augment - Tilastan - Size 7-8, Medial-Right/Lateral Left, 5mm- Intended for...
Waldemar Link GmbH & Co. KG (Mfg Site) began a recall of LinkSymphoKnee Distal Femoral Augment - Tilastan - Size 7-8, Medial-Right/Lateral Left, 5mm- Intended for primary and revision total knee replacement in... on . FDA classified it as Class II and gives the reason as "Fixation screw was inserted through the wrong side of the augment (distal to proximal, as opposed to proximal to distal, which is the correct orientation)". As of , the recall status on record is Open, Classified.
Recall record
- Recall number
- Z-2066-2023
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Waldemar Link GmbH & Co. KG
- Product
- LinkSymphoKnee Distal Femoral Augment - Tilastan - Size 7-8, Medial-Right/Lateral Left, 5mm- Intended for primary and revision total knee replacement in skeletally mature patients. Product Code: 880-307/11
- Product code
- JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer
- Reason for recall
- Fixation screw was inserted through the wrong side of the augment (distal to proximal, as opposed to proximal to distal, which is the correct orientation).
- FDA root cause
- Component change control
- Action
- LinkBio Corporation issued A Product Advisory Notice to consignees holding field inventory on 18-May-2023. The product is provided to the surgeon directly by the representative of the Distributor. Letter states reason for recall, health risk and action to take: Please ensure the surgical team is made aware of this issue before using one of the affected augments. While there is no safety impact due to this issue, there could be some confusion during assembly if the issue is not known in advance. No additional action is required by users beyond reinserting the screw in the correct orientation, and no product return is required at this time. Please contact your LinkBio Sales or Customer Service representative or the contact for any questions: Jennifer Lubrecht Senior Director, Quality Management LinkBio Corporation 69 King Street Dover, NJ 07801 Mobile: +1 862.477.0370 Email: [email protected]
- Quantity in commerce
- 20 units
- Distribution
- US Nationwide distribution in the states of AL, IL, IN, LA, GA, NV, TX.
- Product codes and lots
- UDI-DI: 04026575257379 Lot Number: 2123225
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2066-2023 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2067-2023 | Class II | Ongoing | Voluntary: Firm initiated |