Z-2110-2023Class I

Abbott Amplatzer Steerable Delivery Sheath (ASDS), REF ASDS-14F-075, is a percutaneous catheter (cardiac)

Abbott Medical / initiated 2023-06-12 / Open, Classified / root cause: design

Abbott began a recall of Abbott Amplatzer Steerable Delivery Sheath (ASDS), REF ASDS-14F-075, is a percutaneous catheter (cardiac) on . FDA classified it as Class I and gives the reason as "There is an overall reported incidence rate of 0.77% related to observed or potential air embolism during procedures in which the ASDS product was used". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.

Recall record

Recall number
Z-2110-2023
Classification
Class I
Status
Open, Classified
Initiated
Posted
Recalling firm
Abbott Medical
Product
Abbott Amplatzer Steerable Delivery Sheath (ASDS), REF ASDS-14F-075, is a percutaneous catheter (cardiac).
Product code
DQY Catheter, Percutaneous
Reason for recall
There is an overall reported incidence rate of 0.77% related to observed or potential air embolism during procedures in which the ASDS product was used.
Root cause tag
design
FDA root cause
Device Design
Action
Abbott issued a MEDICAL DEVICE RECALL notice to its consignees beginning on 06/12/2023. The notice explained the problem with the device, the risk, and requested the return of all units. Should you have questions about this issue, please contact your local Abbott Representative or Abbott Support at 1-800-544-1664 (Option 2) (U.S.), 7:00 a.m. - 7:00 p.m. Central Time, Monday through Friday.
Quantity in commerce
675 units
Distribution
Worldwide distribution - US Nationwide and the countries of Slovakia, Czech Republic.
Product codes and lots
UDI/DI 05415067036025, Batch Numbers: 8597228, 8598983, 8601676, 8611443, 8619197, 8627877, 8650955, 8663679, 8795668, 8815423, 8824607, 8830530, 8846835

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2110-2023
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2110-2023Class IOngoingVoluntary: Firm initiated