Z-2110-2023Class I
Abbott Amplatzer Steerable Delivery Sheath (ASDS), REF ASDS-14F-075, is a percutaneous catheter (cardiac)
Abbott Medical / initiated 2023-06-12 / Open, Classified / root cause: design
Abbott began a recall of Abbott Amplatzer Steerable Delivery Sheath (ASDS), REF ASDS-14F-075, is a percutaneous catheter (cardiac) on . FDA classified it as Class I and gives the reason as "There is an overall reported incidence rate of 0.77% related to observed or potential air embolism during procedures in which the ASDS product was used". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-2110-2023
- Classification
- Class I
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Abbott Medical
- Product
- Abbott Amplatzer Steerable Delivery Sheath (ASDS), REF ASDS-14F-075, is a percutaneous catheter (cardiac).
- Product code
- DQY Catheter, Percutaneous
- Reason for recall
- There is an overall reported incidence rate of 0.77% related to observed or potential air embolism during procedures in which the ASDS product was used.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Abbott issued a MEDICAL DEVICE RECALL notice to its consignees beginning on 06/12/2023. The notice explained the problem with the device, the risk, and requested the return of all units. Should you have questions about this issue, please contact your local Abbott Representative or Abbott Support at 1-800-544-1664 (Option 2) (U.S.), 7:00 a.m. - 7:00 p.m. Central Time, Monday through Friday.
- Quantity in commerce
- 675 units
- Distribution
- Worldwide distribution - US Nationwide and the countries of Slovakia, Czech Republic.
- Product codes and lots
- UDI/DI 05415067036025, Batch Numbers: 8597228, 8598983, 8601676, 8611443, 8619197, 8627877, 8650955, 8663679, 8795668, 8815423, 8824607, 8830530, 8846835
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2110-2023 | Class I | Ongoing | Voluntary: Firm initiated |