GE Healthcare Vivid S60N, ultrasound device
GE Vingmed Ultrasound As began a recall of GE Healthcare Vivid S60N, ultrasound device on . FDA classified it as Class II and gives the reason as "GE HealthCare has become aware that certain Vivid ultrasound systems cannot boot up in a timely fashion. If this occurs, it can delay availability of the device in time-critical emergency situations". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2126-2023
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- GE Vingmed Ultrasound AS
- Product
- GE Healthcare Vivid S60N, ultrasound device
- Product code
- IYN System, Imaging, Pulsed Doppler, Ultrasonic
- Reason for recall
- GE HealthCare has become aware that certain Vivid ultrasound systems cannot boot up in a timely fashion. If this occurs, it can delay availability of the device in time-critical emergency situations.
- Root cause tag
- manufacturing process
- FDA root cause
- Process design
- Action
- GE Healthcare sent an Urgent Medical Device Correction notice to its consignees on 05/30/2023 by mail (traceable means). The notice explained the problem with the instrument and provided the following: "You can continue to use your device. Please follow clinical practice guidelines, which include having a backup imaging plan when performing time-critical examinations or image-guided interventions." The devices will be corrected at no charge to the consignee.
- Distribution
- Worldwide distribution.
- Product codes and lots
- Software versions: v203, v204, v205, v206
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2124-2023 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2125-2023 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2126-2023 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2127-2023 | Class II | Ongoing | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.