Z-2128-2023Class II

Abiomed 14Fr Low Profile Introducer Set, Product Number 1000457

Abiomed, Inc. / initiated 2023-05-26 / Open, Classified / root cause: manufacturing process

Abiomed, Inc. began a recall of Abiomed 14Fr Low Profile Introducer Set, Product Number 1000457 on . FDA classified it as Class II and gives the reason as "There was one complaint that the sidearm of the device detached during explant of a pump. When the sidearms are subjected to heat, humidity and articulation, a silicone oil contaminant that was unintentionally...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2128-2023
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Abiomed, Inc.
Product
Abiomed 14Fr Low Profile Introducer Set, Product Number 1000457
Product code
DYB Introducer, Catheter
Reason for recall
There was one complaint that the sidearm of the device detached during explant of a pump. When the sidearms are subjected to heat, humidity and articulation, a silicone oil contaminant that was unintentionally introduced during manufacturing at the joint between the sidearm and the hub may cause the joint to detach.
Root cause tag
manufacturing process
FDA root cause
Under Investigation by firm
Action
An URGENT: VOLUNTARY MEDICAL DEVICE RECALL (REMOVAL) notification letter was delivered to customers on 5/23/23. Customers were also contacted via telephone on 5/26/23. ACTIONS TO BE TAKEN BY THE CUSTOMER/USER: 1. Examine your inventory immediately to determine if you have product subject to this recall (removal). DO NOT USE THE SUBJECT PRODUCTS. 2. Contact your Abiomed local clinical field staff to coordinate the return of the subject products. 3. Review, complete all fields, sign, and return the attached Business Response Form (BRF) on the last page of this letter to the Recall Coordinator identified in this document. 4. Forward this notice to any personnel in your facility who needs to be informed. 5. If any of the product subject to this recall (removal) has been forwarded to another facility, contact that facility immediately and provide a copy of this notice to the relevant personnel. 6. Post a copy of this field safety notice in a visible area for awareness. 7. As with any medical device, adverse reactions or quality problems experienced with the use of this product should be reported to the FDA s MedWatch Adverse Event Reporting Program using the link below: https://www.fda.gov/safety/medwatch-fda-safety-information-and-adverse-event-reportingprogram/reporting-serious-problems-fda If you have questions or concerns regarding this notice, please contact (Shashi Thoutam) directly at +1(734) 262-6255 and/or your local clinical field staff.
Quantity in commerce
37 units
Distribution
US Nationwide distribution in the states of MN, NJ, TN.
Product codes and lots
UDI-DI: 00813502013252; Lot Numbers: 1727033, 1725590

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2128-2023
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2128-2023Class IIOngoingVoluntary: Firm initiated