Z-2226-2023Class II

A901 Communication Manager Application, used in conjunction with the A710 Clinician Programmer application

Medtronic Neuromodulation / initiated 2023-06-21 / Open, Classified / root cause: software

Medtronic Neuromodulation began a recall of A901 Communication Manager Application, used in conjunction with the A710 Clinician Programmer application on . FDA classified it as Class II and gives the reason as "When an update to the A710 Clinician Programmer software application for the Intellis Neurostimulation System was released, the corresponding version of the A901 Communication Manager software application was not...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.

Recall record

Recall number
Z-2226-2023
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Medtronic Neuromodulation
Product
A901 Communication Manager Application, used in conjunction with the A710 Clinician Programmer application..
Product code
LGW Stimulator, Spinal-Cord, Totally Implanted For Pain Relief
Reason for recall
When an update to the A710 Clinician Programmer software application for the Intellis Neurostimulation System was released, the corresponding version of the A901 Communication Manager software application was not available for download.
Root cause tag
software
FDA root cause
Under Investigation by firm
Action
Only one of the two consignees received the notification since one of the consignees independently updated their tablet. On 6/22/2023, the notification dated June 2023 was issued to the consignee via visit by the sales representative. The notification described the issue and provided required actions which included: (1) Update the A901 Comm Manager application to version 1.0.1239 on the clinician programmer (tablet), serial #NPL3011706, using the instructions listed in Appendix A, Updating Clinician Programmer Software; (2) Complete the Customer Confirmation Form enclosed with the letter acknowledging that the consignee has received this information; and (3) Share this notice with all those who need to be aware of the issue within the organization and maintain a copy of it in the records.
Quantity in commerce
2 clinician tablets
Distribution
International distribution to the countries of Israel and Serbia.
Product codes and lots
Serial numbers NPL3011706 and NPL3016791, UDI-DI 643169738058.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2226-2023
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2226-2023Class IIOngoingVoluntary: Firm initiated