REINFORCED INTRODUCER SET, Intra-Aortic Balloon Catheter and Accessories, used to provide counter pulsation...
Datascope Corp. began a recall of REINFORCED INTRODUCER SET, Intra-Aortic Balloon Catheter and Accessories, used to provide counter pulsation therapy in the aorta Product Description Part... on . FDA classified it as Class II and gives the reason as "During IAB catheter insertion, the introducer dilator may fracture at the hub, leaving the body of the introducer dilator housed within the sheath. Risks to health include damage to the femoral artery, descending aorta...". As of , the recall status on record is Open, Classified.
Recall record
- Recall number
- Z-2224-2023
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Datascope Corp.
- Product
- REINFORCED INTRODUCER SET, Intra-Aortic Balloon Catheter and Accessories, used to provide counter pulsation therapy in the aorta Product Description Part Number REINFORCED INTRODUCER SET FOR LINEAR 7.5Fr. & MEGA 7.5Fr. D684-00-0403-05 REINFORCED INTRODUCER SET FOR SENSATION 7Fr. D684-00-0403-06 REINFORCED INTRODUCER SET FOR SENSATION PLUS & MEGA 8Fr. D684-00-0403-10
- Product code
- DYB Introducer, Catheter
- Reason for recall
- During IAB catheter insertion, the introducer dilator may fracture at the hub, leaving the body of the introducer dilator housed within the sheath. Risks to health include damage to the femoral artery, descending aorta, or embolization; surgical intervention and delay of IAB therapy may also result. Firm has received 10 complaints, including 3 serious adverse events and 1 patient death.
- FDA root cause
- Under Investigation by firm
- Action
- Firm notified customers on June 23, 2023 via letter. Customers were made aware of the potential for the device to break during use and the risk of using the device. Customers were given alternate solutions to continue using the IAB Catheter with alternative sheath/dilator combinations. Customers may also return their affected product for a credit. The firm will continue to ship product to customers who choose to continue receiving it. However, product currently being shipped is potentially affected by the same issue. Datascope/Getinge has identified the issue with the introducer dilator and is working with the supplier towards a resolution. If you have any questions, please contact your Datascope/Getinge representative or call Datascope/Getinge Customer Service at 1-888-943-8872, options 2, Monday through Friday, between the hours of 8:00 a.m. and 6:00 p.m. (Eastern Time Zone).
- Quantity in commerce
- 321,609 total kits
- Distribution
- Domestic distribution nationwide. International distribution worldwide.
- Product codes and lots
- Product Description UDI REINFORCED INTRODUCER SET FOR LINEAR 7.5Fr. & MEGA 7.5Fr. 10607567106656 REINFORCED INTRODUCER SET FOR SENSATION 7Fr. 10607567106694 REINFORCED INTRODUCER SET FOR SENSATION PLUS & MEGA 8Fr. 10607567107943
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2216-2023 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2217-2023 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2218-2023 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2219-2023 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2220-2023 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2221-2023 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2222-2023 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2223-2023 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2224-2023 | Class II | Ongoing | Voluntary: Firm initiated |