Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2477-2023Class II

epoc BGEM Crea Test Card with epoc Host SW v3.37.3, epoc NXS SW v4.10.6, Sensor Configuration 41.1

Siemens Healthcare Diagnostics Inc. / initiated 2023-06-01 / Open, Classified

Siemens Healthcare Diagnostics Inc began a recall of epoc BGEM Crea Test Card with epoc Host SW v3.37.3, epoc NXS SW v4.10.6, Sensor Configuration 41.1 on . FDA classified it as Class II and gives the reason as "There is potential for discrepant high glucose results in samples with glucose results on the lower end of the reportable range". As of , the recall status on record is Open, Classified.

Recall record

Recall number
Z-2477-2023
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
epoc BGEM Crea Test Card with epoc Host SW v3.37.3, epoc NXS SW v4.10.6, Sensor Configuration 41.1
Product code
CHL Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph
Reason for recall
There is potential for discrepant high glucose results in samples with glucose results on the lower end of the reportable range.
FDA root cause
Under Investigation by firm
Action
Consignees were notified via an Urgent Medical Device Correction letter beginning June 5, 2023. Consignees were informed of the issue and were provided a workaround and actions to take. The next sensor configuration/software version, sensor configuration 42.1/epoc Host SW v3.38.2 /epoc NXS SW v4.11.11, is available to customers in Siemens Healthineers Document Library or the epoc Live Update Service (eLUS). This next sensor configuration, 42.1, mitigates the issue.
Quantity in commerce
473 ea
Distribution
US distribution to AL, AR, AZ, CA, CO, DC, DE, FL, IA, ID, IL, KS, KY, LA, MI, MN, MO, NC, ND, NH, NY, OH, OK, SC, SD, and TX OUS distribution to Canada, China, Colombia, Finland, India, Mexico, Norway, Sweden, United Arab Emirates, and Vietnam
Product codes and lots
UDI-DI: 00809708072254 Siemens Material Number: 10736382 Lots: 03-23002-60 (expiry 19-Jun-2023) and 03-23056-50 (expiry 12-Aug-2023)

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2477-2023
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2476-2023Class IIOngoingVoluntary: Firm initiated
Z-2477-2023Class IIOngoingVoluntary: Firm initiated
Z-2478-2023Class IIOngoingVoluntary: Firm initiated