Z-2478-2023Class II
epoc BGEM Test Card (BUN & TCO2),Veterinary with epoc Host SW v3.37.3, epoc NXS SW v4.10.6, Sensor...
Siemens Healthcare Diagnostics Inc. / initiated 2023-06-01 / Open, Classified
Siemens Healthcare Diagnostics Inc began a recall of epoc BGEM Test Card (BUN & TCO2),Veterinary with epoc Host SW v3.37.3, epoc NXS SW v4.10.6, Sensor Configuration 41.1 on . FDA classified it as Class II and gives the reason as "There is potential for discrepant high glucose results in samples with glucose results on the lower end of the reportable range". As of , the recall status on record is Open, Classified.
Recall record
- Recall number
- Z-2478-2023
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Siemens Healthcare Diagnostics Inc.
- Product
- epoc BGEM Test Card (BUN & TCO2),Veterinary with epoc Host SW v3.37.3, epoc NXS SW v4.10.6, Sensor Configuration 41.1
- Product code
- CHL Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph
- Reason for recall
- There is potential for discrepant high glucose results in samples with glucose results on the lower end of the reportable range.
- FDA root cause
- Under Investigation by firm
- Action
- Consignees were notified via an Urgent Medical Device Correction letter beginning June 5, 2023. Consignees were informed of the issue and were provided a workaround and actions to take. The next sensor configuration/software version, sensor configuration 42.1/epoc Host SW v3.38.2 /epoc NXS SW v4.11.11, is available to customers in Siemens Healthineers Document Library or the epoc Live Update Service (eLUS). This next sensor configuration, 42.1, mitigates the issue.
- Quantity in commerce
- 348 ea
- Distribution
- US distribution to AL, AR, AZ, CA, CO, DC, DE, FL, IA, ID, IL, KS, KY, LA, MI, MN, MO, NC, ND, NH, NY, OH, OK, SC, SD, and TX OUS distribution to Canada, China, Colombia, Finland, India, Mexico, Norway, Sweden, United Arab Emirates, and Vietnam
- Product codes and lots
- UDI-DI: 00630414243788 Siemens Material Number: 10736516 Lots: 04-23043-51 (expiry 30-Jul-2023)
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2476-2023 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2477-2023 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2478-2023 | Class II | Ongoing | Voluntary: Firm initiated |