QIAstat-Dx Respiratory SARS-CoV-2 Panel -IVD intended for the detection and differentiation of nucleic acid...
Qiagen GmbH began a recall of QIAstat-Dx Respiratory SARS-CoV-2 Panel -IVD intended for the detection and differentiation of nucleic acid from SARS-CoV-2 and the following organism types... on . FDA classified it as Class II and gives the reason as "If cartridges of the affected Serial numbers (SNs) are used, in most cases, the error codes 0xY0003EA and 0xY00045A (where Y is a number between 1 and 4) that lead to a run abortion will occur and no results will be...". As of , the recall status on record is Open, Classified.
Recall record
- Recall number
- Z-2215-2023
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Qiagen
- Product
- QIAstat-Dx Respiratory SARS-CoV-2 Panel -IVD intended for the detection and differentiation of nucleic acid from SARS-CoV-2 and the following organism types and subtypes: Adenovirus, Coronavirus 229E, Coronavirus HKU1, Coronavirus NL63, Coronavirus OC43, SARS-CoV-2, Human Metapneumovirus A+B, Influenza A, Influenza A H1, Influenza A H3, Influenza A H1N1/pdm09, Influenza B, Parainfluenza virus 1, Parainfluenza virus 2, Parainfluenza virus 3, Parainfluenza virus 4, Rhinovirus/Enterovirus, Respiratory Syncytial Virus A+B, Bordetella pertussis, Chlamydophila pneumoniae, and Mycoplasma pneumoniae REF 691223
- Product code
- QJR Reagents, 2019-Novel Coronavirus Nucleic Acid
- Reason for recall
- If cartridges of the affected Serial numbers (SNs) are used, in most cases, the error codes 0xY0003EA and 0xY00045A (where Y is a number between 1 and 4) that lead to a run abortion will occur and no results will be reported
- FDA root cause
- Mixed-up of materials/components
- Action
- Qiagen issued Urgent Medical Device Correction Correction letter via e-mail on 5/9/23. For bounced e-mails, alternative suitable contacts will be identified. Letter states reason for recall, health risk and action to take: Please check if you have remaining stock of LOT 175010712. Do not use the remaining stock of cartridges of LOT 175010712. Dispose of it immediately in accordance to your national and local safety and environmental regulations. Please contact QIAGEN Technical Services for a free-of-charge replacement. Review this notice with your laboratory/medical director. IMPORTANT: Forward this information to all individuals and departments within your organization using the above listed kits. If you are not the end user, please forward this notice to the product end user. Complete the Acknowledgement of Receipt Form attached to this letter by 19. May 2023. Actions taken by QIAGEN QIAGEN was able to identify the root cause of the issue and has implemented immediate actions to ensure that the currently produced cartridges will not cause this issue moving forward. If you have any questions or concerns, please contact your local QIAGEN Technical Services Department through any of the following: Telephone: 800 362 7737 Email: [email protected]
- Quantity in commerce
- 368 kits
- Distribution
- AL, AR, CA, DC, FL, GA, ID, KS, KY, MD, MN, NJ, NY,PA, SC, TX, VA
- Product codes and lots
- GTIN: 14053228038846 LOT Number: 175010712 Affected Serial numbers (SNs) are from 330559516 to 330559783
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2215-2023 | Class II | Ongoing | Voluntary: Firm initiated |