Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2418-2023Class II

Snowden-Pencer MIS DIAMOND-TOUCH DISSECTOR FENESTRATED 16MM JAW 5MM, 32CM, NON RATCHETED, REF SP90-7036...

Medtronic, Inc. / initiated 2023-07-13 / Open, Classified / root cause: manufacturing process

Carefusion 2200 Inc began a recall of Snowden-Pencer MIS DIAMOND-TOUCH DISSECTOR FENESTRATED 16MM JAW 5MM, 32CM, NON RATCHETED, REF SP90-7036; endoscopic surgical device on . FDA classified it as Class II and gives the reason as "Specific lots of the product have the potential to exhibit cracking or breaking of the rotation knob due to the incorrect use of an adhesive in the manufacturing process". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2418-2023
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Medtronic, Inc.
Product
Snowden-Pencer MIS DIAMOND-TOUCH DISSECTOR FENESTRATED 16MM JAW 5MM, 32CM, NON RATCHETED, REF SP90-7036; endoscopic surgical device
Product code
GCJ Laparoscope, General & Plastic Surgery
Reason for recall
Specific lots of the product have the potential to exhibit cracking or breaking of the rotation knob due to the incorrect use of an adhesive in the manufacturing process.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
BD issued an URGENT: Medical Device Product Recall notice to its consignees on 13 July 2023 via FedEx 2-day and email. The notice explained the problem with the device, the risk to the patient, and requested the following actions: Medical facilities: - Discontinue the use of the affected device and destroy it - Share the notice with all users within your facility network - If the product was purchased through a distributor, contact the distributor directly for further instructions. Distributors were directed to Identify all customers within your distribution network that purchased any affected product as defined in this notification. Provide a copy of the attached customer letter to all customers to advise them of this recall notification on BD s behalf. Questions: Phone: 1-844-8BD- LIFE (1-844-823-5433), Say Recall when prompted, M-F 8am - 5pm CT Email: [email protected]
Quantity in commerce
0 units
Distribution
US and Canada
Product codes and lots
UDI/DI 10885403159862, Lot Number B23

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2418-2023
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2415-2023Class IIOngoingVoluntary: Firm initiated
Z-2416-2023Class IIOngoingVoluntary: Firm initiated
Z-2417-2023Class IIOngoingVoluntary: Firm initiated
Z-2418-2023Class IIOngoingVoluntary: Firm initiated
Z-2419-2023Class IIOngoingVoluntary: Firm initiated
Z-2420-2023Class IIOngoingVoluntary: Firm initiated
Z-2421-2023Class IIOngoingVoluntary: Firm initiated
Z-2422-2023Class IIOngoingVoluntary: Firm initiated
Z-2423-2023Class IIOngoingVoluntary: Firm initiated
Z-2424-2023Class IIOngoingVoluntary: Firm initiated
Z-2425-2023Class IIOngoingVoluntary: Firm initiated
Z-2426-2023Class IIOngoingVoluntary: Firm initiated
Z-2427-2023Class IIOngoingVoluntary: Firm initiated
Z-2428-2023Class IIOngoingVoluntary: Firm initiated
Z-2429-2023Class IIOngoingVoluntary: Firm initiated