DYND3000xxP Series - Flexible Laryngeal Mask Airway Packs
MEDLINE INDUSTRIES, LP - Northfield began a recall of DYND3000xxP Series - Flexible Laryngeal Mask Airway Packs on . FDA classified it as Class II and gives the reason as "The case and each (i.e. packet) labels for the DYND3000xxP series do not include an expiration date, however, the lubricating jelly (MDS032280) included in the DYNDP3000xxP series is labeled with an expiration date". As of , the recall status on record is Open, Classified, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-2227-2023
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- MEDLINE INDUSTRIES, LP - Northfield
- Product
- DYND3000xxP Series - Flexible Laryngeal Mask Airway Packs
- Product code
- CAE Airway, Oropharyngeal, Anesthesiology
- Reason for recall
- The case and each (i.e. packet) labels for the DYND3000xxP series do not include an expiration date, however, the lubricating jelly (MDS032280) included in the DYNDP3000xxP series is labeled with an expiration date.
- Root cause tag
- labeling and UDI
- FDA root cause
- Labeling Change Control
- Action
- The consignee recall notification was sent out via first class mail and email on 03/28/2023. The letter instructs the consignee to over-label the Flexible Laryngeal Mask Airway Packs stating to Check the Expiration date of Lubricating Jelly MDS032280 and do not use if Expired . No product is to be returned back to Medline Industries, LP.
- Quantity in commerce
- 9945 ea
- Distribution
- Worldwide distribution - US Nationwide and the country of Bermuda.
- Product codes and lots
- All Lots for series since product launch; Model Number (Unique Device Identifier) -DYND300010P (Each level packaging: 10193489010671; Case level packaging: 20193489010678); DYND300015P (Each level packaging: 10193489010688; Case level packaging: 20193489010685); DYND300020P (Each level packaging: 10193489010695; Case level packaging: 20193489010692); DYND300025P (Each level packaging: 10193489010701 Case level packaging: 20193489010708); DYND300030P (Each level packaging: 10193489010718; Case level packaging: 20193489010715); DYND300040P (Each level packaging: 10193489010725; Case level packaging: 20193489010722); DYND300050P (Each level packaging: 10193489010732; Case level packaging: 20193489010739)