Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2227-2023Class II

DYND3000xxP Series - Flexible Laryngeal Mask Airway Packs

MEDLINE INDUSTRIES, LP - Northfield / initiated 2023-03-28 / Open, Classified / root cause: labeling and UDI

MEDLINE INDUSTRIES, LP - Northfield began a recall of DYND3000xxP Series - Flexible Laryngeal Mask Airway Packs on . FDA classified it as Class II and gives the reason as "The case and each (i.e. packet) labels for the DYND3000xxP series do not include an expiration date, however, the lubricating jelly (MDS032280) included in the DYNDP3000xxP series is labeled with an expiration date". As of , the recall status on record is Open, Classified, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-2227-2023
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
MEDLINE INDUSTRIES, LP - Northfield
Product
DYND3000xxP Series - Flexible Laryngeal Mask Airway Packs
Product code
CAE Airway, Oropharyngeal, Anesthesiology
Reason for recall
The case and each (i.e. packet) labels for the DYND3000xxP series do not include an expiration date, however, the lubricating jelly (MDS032280) included in the DYNDP3000xxP series is labeled with an expiration date.
Root cause tag
labeling and UDI
FDA root cause
Labeling Change Control
Action
The consignee recall notification was sent out via first class mail and email on 03/28/2023. The letter instructs the consignee to over-label the Flexible Laryngeal Mask Airway Packs stating to Check the Expiration date of Lubricating Jelly MDS032280 and do not use if Expired . No product is to be returned back to Medline Industries, LP.
Quantity in commerce
9945 ea
Distribution
Worldwide distribution - US Nationwide and the country of Bermuda.
Product codes and lots
All Lots for series since product launch; Model Number (Unique Device Identifier) -DYND300010P (Each level packaging: 10193489010671; Case level packaging: 20193489010678); DYND300015P (Each level packaging: 10193489010688; Case level packaging: 20193489010685); DYND300020P (Each level packaging: 10193489010695; Case level packaging: 20193489010692); DYND300025P (Each level packaging: 10193489010701 Case level packaging: 20193489010708); DYND300030P (Each level packaging: 10193489010718; Case level packaging: 20193489010715); DYND300040P (Each level packaging: 10193489010725; Case level packaging: 20193489010722); DYND300050P (Each level packaging: 10193489010732; Case level packaging: 20193489010739)

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2227-2023
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2227-2023Class IIOngoingVoluntary: Firm initiated