Z-2240-2023Class II

Cios Select VA 21 with FD- A Mobile X-ray system intended for use in Operating room, Traumatology, Endoscopy...

Siemens Medical Solutions USA, Inc. / initiated 2023-06-05 / Open, Classified / root cause: manufacturing process

Siemens Medical Solutions USA, Inc began a recall of Cios Select VA 21 with FD- A Mobile X-ray system intended for use in Operating room, Traumatology, Endoscopy, Intensive Care Station, Pediatrics, Ambulatory... on . FDA classified it as Class II and gives the reason as "Incorrect glue was applied on the plexiglass sheet on the single tank cover assembly could potentially fall off and the ingress of liquid may occur, the risk of electric shock exists (e.g., ingress of a large amount of...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2240-2023
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
Cios Select VA 21 with FD- A Mobile X-ray system intended for use in Operating room, Traumatology, Endoscopy, Intensive Care Station, Pediatrics, Ambulatory patient care and in Veterinary Medicine. Material #11515088
Product code
OWB Interventional Fluoroscopic X-Ray System
Reason for recall
Incorrect glue was applied on the plexiglass sheet on the single tank cover assembly could potentially fall off and the ingress of liquid may occur, the risk of electric shock exists (e.g., ingress of a large amount of liquid following the plexiglass dropping off, at the same time the user or operator happens to touch the liquid on the cover which is connected to the internal live parts).
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Siemens issued Customer Safety Advisory Notice via AX015/23/S on 6/6/23. Letter states reason for recall, health risk and action to take: Siemens will correct the hardware error via Update Instruction AX012/23/S. This measure will replace the current single tank cover of all affected systems. Our service organization will contact you shortly to arrange a date to perform this corrective action. Please feel free to contact our service organization for an earlier appointment at 1-800-888-7436. Please review, understand, and confirm acknowledgement of the information and instructions outlined in this Urgent Safety Advisory Notice. By signing this document, you are acknowledging that you have read and understand the content therein
Quantity in commerce
9 units
Distribution
Worldwide - US Nationwide distribution in the states of FL, GA, KS, MN, NC, NY and the countries of Austria, Bangladesh, Brazil, Canada, China, Croatia, El Salvador, France, Germany, Indonesia, Italy, Japan, Jordan, Lebanon, Luxembourg, Mauritius, Mexico, Monaco, Peru, Philippines, Poland, Romania, Slovakia, South Africa, Spain, Switzerland, Taiwan, Ukraine, United Kingdom.
Product codes and lots
UDI-DI :04056869148243 US S/N: 30001 30002 30008 30013 30029 30031 30032 30033 30034

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2240-2023
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2240-2023Class IIOngoingVoluntary: Firm initiated