Z-2248-2023Class I

Carina Sub-Acute Care Ventilator

Draeger, Inc. / initiated 2023-07-12 / Open, Classified / root cause: manufacturing process

Draeger Medical, Inc. began a recall of Carina Sub-Acute Care Ventilator on . FDA classified it as Class I and gives the reason as "Presence of polyether polyurethane (PE-PUR) and emission of 1,3-Dichloropropan-2-ol that exceed the acceptable uptake level during continuous use (>30 days) in pediatric patients". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2248-2023
Classification
Class I
Status
Open, Classified
Initiated
Posted
Recalling firm
Draeger, Inc.
Product
Carina Sub-Acute Care Ventilator
Product code
CBK Ventilator, Continuous, Facility Use
Reason for recall
Presence of polyether polyurethane (PE-PUR) and emission of 1,3-Dichloropropan-2-ol that exceed the acceptable uptake level during continuous use (>30 days) in pediatric patients.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
The firm, Drager, sent an "Urgent Medical Device Recall" to customers by mail on 07/12/2023. The medical device recall notification instructs customers on how to utilize the Carina Ventilator until units can be corrected in the field by way of replacement of the blower to remove the presence of PE-PUR foam. Customers may continue to use the device until it has been modified provided that the set minute volume exceeds 3.6 l/min, only adult patients are ventilated, and only a Draeger ventilation hose with leak valve is used. The firm states that a new designed blower cover will be available in the early part of the fourth quarter of 2023 and Draeger will facilitate field service free of charge. Customers are to ensure all users and maintenance staff that utilize the recalled device are made aware of the recall notice. Customers with any questions regarding the operation of the Carina Ventilator are to contact Technical Support between 8:00 AM and 8:00 PM EST at 1-800-437-2437 (press 2, then 2, then 2); customers with any questions about the recall notice are to contact Michael Kelhart from 8:00 AM to 4:30 PM EST at 267-664-1131 or by email at [email protected]. The Medical Device Recall Response form is to be returned by fax 215-372-2940 or email: [email protected] to Draeger upon completion.
Quantity in commerce
11,621 units
Distribution
US Nationwide Distribution.
Product codes and lots
Part No. 5704110; UDI-DI 04048675398516; All Serial No.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2248-2023
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2248-2023Class IOngoingVoluntary: Firm initiated