Plum 360 Infusion System, List Numbers: 300100405, 300100409, 300100413, 300100414, 300100415, 300100418...
ICU Medical Inc began a recall of Plum 360 Infusion System, List Numbers: 300100405, 300100409, 300100413, 300100414, 300100415, 300100418, 300100451, 300100452, 300100453, 300100481... on . FDA classified it as Class II and gives the reason as "Due to a manufacturing defect of a supplier provided component, there is a potential that the audible signal for an alarm may not sound under certain conditions". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2430-2023
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Abbott Medical
- Product
- Plum 360 Infusion System, List Numbers: 300100405, 300100409, 300100413, 300100414, 300100415, 300100418, 300100451, 300100452, 300100453, 300100481, 300100483, 300100613, 300101013, 300101113, 300101114, 300101313, 300101379, 300102213, 300102713, 300102913, 300103413, 300103414, 300104213, 300105413, 300105713, 300106313, 300106513, 300106581, 300107813, 300107814, 300108513, 300108813, 300108814, 300109913; infusion pump
- Product code
- FRN Pump, Infusion
- Reason for recall
- Due to a manufacturing defect of a supplier provided component, there is a potential that the audible signal for an alarm may not sound under certain conditions.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- ICU Medical issued an URGENT: MEDICAL DEVICE CORRECTION notice on 07/13/2023 via traceable mailing. The notice explained the issue, potential risk, and requested the following: Actions for Clinical Users: If your Plum 360 pump does not power on or indicates an Audio Alarm failure, please remove the pump from use and send the pump to Biomedical Engineering. Actions for Biomedical Engineering: If you have Plum 360 pumps with Audio Alarm failures, please contact ICU Medical. 1. Locate all affected pumps in your possession and ensure all users or potential users of these devices are immediately made aware of this notification and required actions. 2. Complete and return the attached Response Form to [email protected] within ten days of receipt to acknowledge your understanding of this notification. 3. DISTRIBUTORS: If you have distributed affected products to your customers, please immediately forward this notice to them. Request that they complete the response form and return it to [email protected]. A similar notice was distributed specifically to consignees of instruments that were serviced which lists the specific serial numbers. The issue, potential risk, and actions are the same as the general notice.
- Quantity in commerce
- 93135 devices
- Distribution
- Worldwide
- Product codes and lots
- UDI/DI M335300101, All Plum 360 pumps manufactured between July 2020 and December 2021
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2430-2023 | Class II | Ongoing | Voluntary: Firm initiated |