Tracheal Intubation Fiberscope OLYMPUS LF-P- To be used with an Olympus Light Source, documentation...
Olympus Corporation of the Americas began a recall of Tracheal Intubation Fiberscope OLYMPUS LF-P- To be used with an Olympus Light Source, documentation equipment, display monitor and other ancillary equipment... on . FDA classified it as Class II and gives the reason as "Revised OER-Pro, OER-Elite, OER-Mini endoscope reprocessor labeling- the LF-V and LF-P are no longer compatible endoscopes for reprocessing in the OERs. The addendum highlights the following changes: 1. Removal of LF-V...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-2341-2023
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- OLYMPUS Corporation of the Americas
- Product
- Tracheal Intubation Fiberscope OLYMPUS LF-P- To be used with an Olympus Light Source, documentation equipment, display monitor and other ancillary equipment for management of Endotracheal Tube
- Product code
- EOQ Bronchoscope (Flexible Or Rigid)
- Reason for recall
- Revised OER-Pro, OER-Elite, OER-Mini endoscope reprocessor labeling- the LF-V and LF-P are no longer compatible endoscopes for reprocessing in the OERs. The addendum highlights the following changes: 1. Removal of LF-V and LF-P as compatible with the OER 2. Change conditions for ETO gas sterilization from 12% EtO to 100% EtO 3. Removing the reusable cleaning brush for the manual cleaning. Single Use brushes remain compatible
- Root cause tag
- labeling and UDI
- FDA root cause
- Labeling Change Control
- Action
- Olympus issued two Urgent: Medical Device Corrective Action letters: (1) VISERA TRACHEAL INTUBATION VIDEOSCOPE OLYMPUS LF TYPE V (LF-V) TRACHEAL INTUBATION FIBERSCOPE OLYMPUS LF-P (LF-P) (2)ENDOSCOPE REPROCESSOR Affected Models: OER-ProTM, OER-MiniTM and OER-EliteTM. Letters states reason for recall, health risk and action to take: Our records indicate that your facility has purchased one or more LF-V and LF-P scopes. (1)Olympus requests you to take the following actions: 1. Ensure all personnel are completely knowledgeable and thoroughly trained on the changes to the reprocessing instructions described in this letter. 2. If you are using OER for high level disinfection of your LF-V and LF-P, please transition to another compatible method of reprocessing as soon as possible. 3. Locate and replace any on-hand copies of the LF-V and LF-P Instructions for Use with the updated version. Obtain a copy of the updated IFU by accessing our Olympus Connect website: How to access the Instructions for Use a) Log in to www.OlympusConnect.com b) Select Product Support , then Instruction Manuals c) Navigate to the document entitled LF-V Instruction Manual and LF-P Instruction Manual If you require additional information about the updated Instruction for use, or need assistance with accessing Olympus Connect, please contact our Technical Assistance Center (TAC) at 1-800-848-9024 (option 1). The TAC department can assist you in answering questions on the new Instructions Manual and new steps. 4. If a Scope ID tag (MAJ-1545) is used with your LF-V and LF-P, please dispose of your MAJ-1545 since the LF-V and LF-P cannot be used with OERs. 5. Complete the enclosed response form and return to our recall partner, Sedgwick, via Email ([email protected]) or Fax (866-842-0470). For any questions about the acknowledgement form, please call the Sedgwick team at 888-912-7350. (2)Our records indicate that your facility has purchased one or more of the afore mentioned endoscope
- Quantity in commerce
- 686 units
- Distribution
- Nationwide
- Product codes and lots
- UDI-DI: 04953170051098 All serial numbers
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2340-2023 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2341-2023 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2342-2023 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2343-2023 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2344-2023 | Class II | Ongoing | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.