Fusion Lithotripsy Extraction Basket, Model Number FS-LXB-2X4 G48277; Fusion Lithotripsy Extraction Baskets...
Wilson-Cook Medical Inc. began a recall of Fusion Lithotripsy Extraction Basket, Model Number FS-LXB-2X4 G48277; Fusion Lithotripsy Extraction Baskets are intended to be used for endoscopic removal of... on . FDA classified it as Class II and gives the reason as "Non-conforming product may appear in the field as inability to advance or retract the basket, or the drive wire breaking or separating from handle (with or without mechanical lithotripsy)". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2252-2023
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Wilson-Cook Medical Inc.
- Product
- Fusion Lithotripsy Extraction Basket, Model Number FS-LXB-2X4 G48277; Fusion Lithotripsy Extraction Baskets are intended to be used for endoscopic removal of biliary stones and foreign bodies.
- Product code
- FFL Dislodger, Stone, Basket, Ureteral, Metal
- Reason for recall
- Non-conforming product may appear in the field as inability to advance or retract the basket, or the drive wire breaking or separating from handle (with or without mechanical lithotripsy).
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Cook Medical issued an URGENT: MEDICAL DEVICE REMOVAL notice to their consignees on 07/14/2023 by email or priority mail. The notice explained the issue, the risk, and requested the consignee cease use and return the affected product.
- Quantity in commerce
- 393 units
- Distribution
- Worldwide distribution.
- Product codes and lots
- UDI/DI 00827002482777, 10827002482774, Lot Numbers: W4589620, W4589141, W4596704, W4596704, W4589141, W4715238, W4596704, W4715238, W4715238, W4646708, W4644270, W4596704, W4589141, W4589141, W4715238, W4646708, W4715238, W4660182, W4596704, W4660182, W4660182, W4589141, W4596704, W4660182, W4715242, W4644270, W4589620, W4664573, W4646708, W4660182, W4589141, W4589620, W4644270, W4660182, W4660182, W4589620, W4644270, W4596704, W4644270, W4664573, W4644270, W4644270, W4646708, W4715238, W4664573, W4608745, W4670102, W4697072, W4697068, W4697072, W4696416, W4660182, W4620539, W4620539, W4620539, W4651764, W4651764, W4670097, W4664041, W4661895, W4647284, W4589621, W4693149, W4645756, W4594582, W4596701, W4612933, W4647284, W4664041, W4612933, W4643861, W4645763, W4594582, W4589621, W4635967, W4631138, W4644271, W4645308, W4645755, W4645756, W4647284, W4631138, W4596701, W4597312, W4612932, W4664041, W4594582, W4594582, W4597312, W4645755, W4661895, W4645763, W4571161, W4608304, W4610258, W4612932, W4612933, W4635967, W4645755, W4645755, W4693149, W4610258, W4596701, W4597312, W4645308, W4647284, W4635967, W4645308, W4693149, W4645763, W4571161, W4597312, W4612933, W4631138, W4631932, W4631932, W4571161, W4662353, W4635967, W4608304, W4612932, W4661895, W4645756, W4635967, W4635967, W4631932, W4635967, W4612933, W4631138, W4645308, W4596701, W4608304, W4662353, W4693149, W4589621, W4612933, W4645755, W4647284, W4662353, W4610258, W4643861, W4644271, W4644271, W4645308, W4645763, W4596701, W4608304, W4608304, W4612932, W4619650, W4622436, W4631932, W4635967, W4631138, W4594582, W4643861, W4631932, W4645756, W4662353, W4664041, W4594582, W4589621, W4664041, W4662353, W4612933, W4612933, W4631138, W4612933, W4631138, W4596704, W4589141
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2252-2023 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2253-2023 | Class II | Ongoing | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY