Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2253-2023Class II

Fusion Lithotripsy Extraction Basket, Model Number FS-LXB-3X6 G48278; Fusion Lithotripsy Extraction Baskets...

Wilson-Cook Medical Inc. / initiated 2023-06-22 / Open, Classified / root cause: manufacturing process

Wilson-Cook Medical Inc. began a recall of Fusion Lithotripsy Extraction Basket, Model Number FS-LXB-3X6 G48278; Fusion Lithotripsy Extraction Baskets are intended to be used for endoscopic removal of... on . FDA classified it as Class II and gives the reason as "Non-conforming product may appear in the field as inability to advance or retract the basket, or the drive wire breaking or separating from handle (with or without mechanical lithotripsy)". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2253-2023
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Wilson-Cook Medical Inc.
Product
Fusion Lithotripsy Extraction Basket, Model Number FS-LXB-3X6 G48278; Fusion Lithotripsy Extraction Baskets are intended to be used for endoscopic removal of biliary stones and foreign bodies.
Product code
FFL Dislodger, Stone, Basket, Ureteral, Metal
Reason for recall
Non-conforming product may appear in the field as inability to advance or retract the basket, or the drive wire breaking or separating from handle (with or without mechanical lithotripsy).
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Cook Medical issued an URGENT: MEDICAL DEVICE REMOVAL notice to their consignees on 07/14/2023 by email or priority mail. The notice explained the issue, the risk, and requested the consignee cease use and return the affected product.
Quantity in commerce
402 units
Distribution
Worldwide distribution.
Product codes and lots
UDI/DI 00827002482784, 10827002482781, Lot Numbers: W4631139, W4702268, W4631139, W4631139, W4693489, W4693489, W4576227, W4702262, W4693489, W4616511, W4693489, W4612934, W4609498, W4693489, W4609498, W4576227, W4631537, W4644272, W4693489, W4702262, W4576227, W4616511, W4612934, W4576227, W4631537, W4644272, W4631537, W4616511, W4702262, W4576227, W4612934, W4576227, W4612934, W4612934, W4612934, W4693489, W4631537, W4644272, W4609498, W4609498, W4702262, W4576227, W4612934, W4576227, W4631537, W4697764, W4576227, W4616511, W4631537, W4697764, W4644272, W4697764, W4702262, W4616511, W4697764, W4702262, W4660177, W4616520, W4610273, W4616520, W4616520, W4635963, W4622445, W4662350, W4645764, W4616520, W4625161, W4701899, W4622445, W4647285, W4647285, W4647285, W4635963, W4645764, W4663702, W4701899, W4645764, W4554892, W4663702, W4701899, W4702270, W4645764, W4616520, W4622445, W4645764, W4660177, W4645764, W4701897, W4702270, W4662350, W4663702, W4662350, W4610273, W4554892, W4664043, W4589622, W4590893, W4590893, W4590893, W4594583, W4594583, W4594583, W4594583, W4594583, W4619627, W4619627, W4619627, W4619651, W4625160, W4625160, W4625160, W4631933, W4647697, W4647701, W4651755, W4651755, W4651765, W4651765, W4651765, W4664574, W4664574, W4660177, W4645764, W4622445, W4610273, W4610273, W4616520, W4625161, W4701899, W4554892, W4554892, W4622445, W4664043, W4701899, W4625161, W4647285, W4554892, W4647285, W4622445, W4702270, W4554892, W4554892, W4554892, W4635963, W4701897, W4616520, W4631139, W4631139, W4631139, W4693489, W4631139, W4644272, W4697073, W4697252, W4631535, W4616511, W4621105, W4609498, W4616511, W4609498, W4554892, W4610273, W4664042, W4693489, W4625161, W4662350, W4610273, W4635963, W4635963, W4647285, W4664043, W4625161, W4660177, W4663702, W4664043, W4701897, W4701899, W4702270, W4662350, W4662350, W4663702, W4664043, W4701897, W4609498, W4664042, W4663702, W4697764, W4664042, W4645764, W4635963, W4702262

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2253-2023
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2252-2023Class IIOngoingVoluntary: Firm initiated
Z-2253-2023Class IIOngoingVoluntary: Firm initiated