Z-2467-2023Class II
Evis Exera III Duodenovideoscope Olympus Tjf-Q190v
OLYMPUS Corporation of the Americas / initiated 2023-07-27 / Open, Classified
Olympus Corporation of the Americas began a recall of EVIS EXERA III DUODENOVIDEOSCOPE OLYMPUS TJF-Q190V on . FDA classified it as Class II and gives the reason as "Recent reports of patient infection". As of , the recall status on record is Open, Classified.
Recall record
- Recall number
- Z-2467-2023
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- OLYMPUS Corporation of the Americas
- Product
- Evis Exera III Duodenovideoscope Olympus Tjf-Q190v
- Product code
- FDT Duodenoscope And Accessories, Flexible/Rigid
- Reason for recall
- Recent reports of patient infection.
- FDA root cause
- Under Investigation by firm
- Action
- Consignees were sent an Urgent Medical Corrective Action notification dated 7/27/2023. This action is being initiated due to the firm becoming aware of and FDA inquiring about recent reports of infections and positive cultures. Olympus is reminding users to closely follow reprocessing instructions, including periodic inspections by Olympus Service, and informing users about an updated manual intended to improve comprehension of existing reprocessing steps. Additional enhancements to the manual include the use of reprocessing timelines and simplified text/visualization techniques. The manual is available on the customer web portal and paper copies, along with wall charts, are available to be mailed to your facility. The notification also emphasises the required inspection of the forceps elevator mechanism, to be performed once a year or every 100 reprocessing cycles, and encourages consignees to consider incorporating voluntary routine microbiologic surveillance of endoscops as part of their quality improvement/assurance efforts. Consignees are asked to complete and return the provided response form to Sedgwick via email at [email protected] or by fax to 888-208-4588. Questions can be directed by phone to 866-912-9552. Customers can also contact Ashley Mitch at Olympus by phone at 484-553-1029 or by email at [email protected] with any additional concerns.
- Quantity in commerce
- 6,426 units
- Distribution
- US Nationwide distribution.
- Product codes and lots
- Model No. TJF-Q190V; UDI-DI: 04953170405563; All Serial No.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2467-2023 | Class II | Ongoing | Voluntary: Firm initiated |