Z-2256-2023Class II

The following Pro.specta Emission Computed Tomography Systems that were manufactured with an SLD Look Ahead...

Siemens Medical Solutions USA, Inc. / initiated 2023-07-12 / Open, Classified / root cause: design

Siemens Medical Solutions USA, Inc. began a recall of The following Pro.specta Emission Computed Tomography Systems that were manufactured with an SLD Look Ahead Optical Sensor Assembly: (a) Symbia Pro.specta Q3... on . FDA classified it as Class II and gives the reason as "During a whole-body planar scan with auto-contour, there is a possibility the short-linear drive (SLD) look-ahead sensors may trigger earlier than expected and may result in detector 1 positioning further from the...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.

Recall record

Recall number
Z-2256-2023
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
The following Pro.specta Emission Computed Tomography Systems that were manufactured with an SLD Look Ahead Optical Sensor Assembly: (a) Symbia Pro.specta Q3, Model 11364751; (b) Symbia Pro.specta X3, Model 11364752; (c) Symbia Pro.specta X7, Model 11364753.
Product code
KPS System, Tomography, Computed, Emission
Reason for recall
During a whole-body planar scan with auto-contour, there is a possibility the short-linear drive (SLD) look-ahead sensors may trigger earlier than expected and may result in detector 1 positioning further from the patient than needed leading to a reduction in image resolution.
Root cause tag
design
FDA root cause
Device Design
Action
The firm issued undated letters flagged Customer Safety Advisory Notice CAN 001-2023 on 7/12/2023 via email, certified mail, or, if necessary, hand delivery. The letter explains the issue, when the malfunction occurs, the potential risks, and how the consignee can help to avoid the potential risk of the issue. The letter informs the consignee that Siemens Healthineers is working to address the issue through a scheduled service visit and the consignee's local service organization will begin contacting the consignees in the 4th quarter of 2023 to schedule this work. In the meantime, the notice is to be placed in the Symbia Pro.specta Operator Manual and disseminated to all operators of the scanner. No response form was enclosed. The recalling firm uses an SAP system for tracking purposes. For the letters that are delivered via email, the consignee verifies receipt through an adobe signature. For the letters delivered via certified mail, there is a return receipt.
Quantity in commerce
108 systems
Distribution
Distribution was made to ID, KS, MA, NH, OH, TX, and VA. There was no government/military distribution. Foreign distribution was made to Argentina, Australia, Austria, Belgium, Canada, Colombia, Denmark, Finland, France, Germany, Hong Kong, Israel, Italy, Japan, Netherlands, Portugal, Singapore, Switzerland, Turkey, United Arab Emirates, and United Kingdom.
Product codes and lots
Software version VA20A. (a) Serial numbers 100201, 100202, 100203, 100204, 100205, 100206, 100207, 100209, 100208, 100210, 100212, 100211, 100213, 100215, 100217, 100219, 100218, 100220, 100221, and 100224; UDI-DI 4056869264745. Additional serial numbers as of 09/07/2023: 100223, 100225, 100227, 100228, 100237; Additional serial numbers as of 10/17/2023: 100214, 100229; Additional serial numbers as of 12/12/2023: 100232, 100231, 100234; Additional serial number as of 01/15/2024:100226. (b) Serial numbers 100101, 100102, 100201, 100203, 100204, 100207, 100206, 100208, 100205, 100209, 100210, 100212, 100213, 100214, 100215, 100216, 100218, 100217, 100219, 100220, 100221, 100223, 100224, 100222, 100226, 100228, 100227, 100225, 100229, 100231, 100232, 100233, 100235, 100234, 100237, 100239, 100241, 100240, 100236, 100238, 100242, 100243, 100247, 100244, 100250, 100251, 100252, and 100255; UDI-DI 4056869264752; Additional serial numbers as of 09/07/2023: 100245, 100256, 100257, 100258, 100261, 100262, 100265, 100273; Additional serial numbers as of 10/17/2023: 100230, 100266; Additional serial numbers as of 12/12/2023: 100269; Additional serial numbers as of 01/15/2024: 100248, 100254, 100260, 100271, 100272. (c) Serial numbers 100101, 100201, 100203, 100202, 100205, and 100206; UDI-DI 4056869264769; Additional serial numbers as of 09/07/2023: 100208, 100210; Additional serial numbers as of 12/12/2023: 100209

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2256-2023
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2256-2023Class IIOngoingVoluntary: Firm initiated