Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2410-2023Class II

Phaco Standalone Tips

Alcon Laboratories / initiated 2023-06-28 / Open, Classified / root cause: manufacturing process

Alcon Research, LTD. began a recall of Phaco Standalone Tips on . FDA classified it as Class II and gives the reason as "Reports of patient injury during use in procedures due to potential for plastic debris from over-tightening with included plastic wrench". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2410-2023
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Alcon Laboratories
Product
Phaco Standalone Tips
Product code
HQC Unit, Phacofragmentation
Reason for recall
Reports of patient injury during use in procedures due to potential for plastic debris from over-tightening with included plastic wrench.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Customers were sent a Voluntary Medical Device Correction Notification by mail and dated 06/29/2023. Customers are asked to follow the risk mitigation precautions detailed in the notice to avoid the instance of overtightening the Phaco Tip and generating plastic particles. Additionally, customers are asked to forward the recall notification to all departments or organizations that use recalled devices. Customers are asked to complete and return the provided Response Form by fax at 817-302-4337 or email at [email protected]. Customers who have experienced any adverse events or quality related issues are to contact alcon via email at [email protected] or by phone at 1-800-757-9780, Option 5.
Quantity in commerce
444,372 units
Distribution
Worldwide distribution - US Nationwide and the countries of Aland Islands, Albania, Algeria, Andorran, Argentina, Armenia, Aruba, Australia, Austria, Azerbaijan, Bahamas, Bahrain, Bangladesh, Barbados, Belarus, Belgium, Bermuda, Botswana, Brazil, Brunei Daruss., Bulgaria, Burma, Cambodia, Canada, Cayman Islands, Chile, China, Colombia, Costa Rica, Croatia, Cura¿ao, Cyprus, Czech Republic, Denmark, Dominican Rep., Ecuador, Egypt, El Salvador, Finland, France, French Polynesia, French Guayana, Georgia, Germany, Ghana, Greece, Guadeloupe, Guernsey, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Isle of Man, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Kenya, Kosovo, Kuwait, Kyrgyzstan, Latvia, Lebanon, Libya, Luxembourg, Macau, Malaysia, Malta, Martinique, Mauritius, Mexico, Moldova, Mongolia, Montenegro, Morocco, Namibia, Netherlands, New Caledonia, New Zealand, Nicaragua, North Macedonia, Norway, Oman, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Puerto Rico, Qatar, Reunion, Romania, Russian Fed., Saint Martin, San Marino, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Syria, Taiwan, Thailand, Trinidad, Tobago, Tunisia, Turkey, Ukraine, United Kingdom, Uruguay, USA, USVI, Utd. Arab Emir., Uzbekistan, Vietnam, and Yemen.
Product codes and lots
Catalog No. 30KTS, 30RTS, 8065740806, 8065740837, 8065750261, 8065750263, 8065750264, 8065750852, 8065750853, 8065751176, 8065751177, 8065751178, 8065752065, 8065752066, 8065790020, 8065790022, 8065790023; Product Code 100000692, 100000691, 100002823, 100002787, 100058543, 100004407, 100004408, 100006953, 100006952, 100006224, 100006225, 100006226, 100009104, 100009105, 100000696, 100000698, 100000708; UDI-DI: 380657403042, 380657403028, 380657408061, 380657408375, 380657502615, 380657502639, 380657502646, 380657508525, 380657508532, 380657511761, 380657511778, 380657511785, 380657520657, 380657520664, 380657900206, 380657900220, 380657900237; Lots 152A2F, 14T8U8, 14V9D7, 14WN7D, 14WNFC, 14YJJD, 15118L, 15KCLD, 14DNN3, 14XP2V, 154DY2, 155L1M, 155LD1, 15EM4L, 15FU4D, 14HA12, 14LH5X, 14LHL4, 14N03H, 14N0A6, 14PLY2, 14PM6Y, 14T8UF, 14T95K, 14UL3F, 14UY8J, 14V94T, 14V9DD, 14YJ8N, 14YJJN, 15075D, 1511AL, 155LD3, 15AYD1, 15AYD3, 15AYD4, 14C5L1, 14V8YF, 1510VH, 153KHX, 15AYF7, 15AYF8, 15CP8C, 151133, 14N03U, 14N0A7, 14PLY1, 14PM6X, 14V9DM, 14YJ93, 14YJJP, 153KR5, 154E88, 15AYD6, 15AYD7, 14C5UK, 14V953, 14YJ94, 1507DP, 153K87, 15AYD8, 15DMRM, 14D015, 14D0J0, 14DYMF, 14HA15, 14LH5W, 14LHL3, 14N03D, 14N0A3, 14PLY0, 14PM6W, 14T8UE, 14T95J, 14UL3C, 14UY8F, 14V94R, 14V9DC, 14WN7N, 14WNFN, 14XNWP, 14XP3Y, 14YJ8D, 14YJJK, 15075M, 1507DM, 15112P, 1511AH, 152A2N, 152ADL, 153K86, 153KR1, 154DY0, 155L2E, 155LD0, 156H57, 15AY6P, 15AY70, 15AYE9, 15AYEA, 15AYEC, 15AYED, 15AYEE, 15AYEF, 15CP8J, 15CPAD, 154E87, 14H9TN, 14HE56, 14LH5U, 14LHL1, 14N03C, 14N0A2, 14PLXY, 14PM6V, 14T8UC, 14T8UD, 14T95F, 14UL39, 14UL3A, 14UY8D, 14UY8E, 14V94M, 14V94N, 14V94P, 14V9D8, 14V9D9, 14V9DA, 14WN7L, 14WN7M, 14WNFL, 14WNFM, 14XNWM, 14XNWN, 14XP3W, 14XP3X, 14YJ89, 14YJ8C, 14YJJH, 14YJJJ, 15075E, 15075F, 1507DK, 1507DL, 15112M, 15112N, 1511AE, 1511AF, 152A2L, 152A2M, 152ADJ, 152ADK, 153K84, 153K85, 153KPW, 154DXX, 154E85, 155L2C, 155L2D, 155LCX, 155LCY, 15AY6M, 15AY6X, 15AYDP, 15AYDR, 15AYDT, 15AYE0, 15AYE2, 15AYE3, 15AYE4, 15AYE5, 15AYE6, 15AYE7, 15AYE8, 15CP8F, 15CP8H, 15CPAA, 15EM3D, 14MV1X, 14MV1Y, 14MV20, 14N03V, 14N0A8, 14PLXX, 14PM6U, 14T8UA, 14T95E, 14WN7H, 14WNFH, 14XNWJ, 14XP3U, 14YJ86, 14YJJE, 15075P, 1507DR, 15112J, 1511AA, 152A2K, 152ADF, 153KPU, 154DXV, 154E83, 155L28, 156H54, 156H6X, 15AY6U, 15AY73, 15AYDA, 15AYDD, 15AYDE, 15AYDF, 15AYDJ, 15CPAJ, 15DN0K, 14C5KL, 14LH5T, 14LHL0, 14WN7J, 14WNFJ, 14XNWK, 14YJ88, 15112K, 1511AC, 1589D4, 15AY6V, 146X4J, 14C5UL, 14UL38, 14UY8C, 14WN7K, 14WNFK, 14XNWL, 14XP3V, 14YJJF, 15112L, 15AYDL, 15DMRP, 14HA16, 14LHD7, 14N03E, 14N0A4, 14PLRJ, 14PM2C, 14T8LX, 14T90K, 14UL3D, 14UY8H, 14V8YD, 14V96T, 14WN7P, 14WN7R, 14WNFP, 14WNFR, 14XNWH, 14XP3T, 14YJ8F, 14YJJL, 15075R, 1507DT, 15112R, 1511AJ, 152A2E, 152ADA, 153K12, 153KHV, 154DR1, 154E0U, 155L2H, 155L2J, 155L98, 156H58, 156H73, 156H88, 15AY6R, 15AY71, 15AYEM, 14PLRK, 14PM2D, 14T8LY, 14T90L, 14UKXE, 14UY4A, 14V8YE, 14V96U, 14WN7T, 14WNFT, 14XNWF, 14XP3R, 14YJ8M, 14YJJM, 15075N, 1507DN, 15112T, 1529WJ, 152A4P, 153K13, 153KHW, 154DR2, 154E0V, 15AYEU, 15AYEW, 15AYEX, 15AYEY, 15AYF0, 15AYF1, 15AYF2, 15AYF3, 15AYF4, 15AYF5, 15AYF6, 15EM34, 150752, 14LH5P, 14LHKX, 14T95D, 14WN7E, 14WNFE, 14XP4P, 150752, 1507DF, 15110P, 15118M, 152ADD, 153KPN, 154DXU, 154E82, 155LCU, 15AYCU, 15AYCV, 15AYCW, 15E58E, 146WR7, 146X46, 14D0HP, 14UL37, 14UY8A, 14WN7F, 14WNFF, 14XP4R, 1507DH, 15110R, 15118N, 152A2J, 152ADE, 15AYCY, 15DMVX, 15AYD0.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2410-2023
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2406-2023Class IIOngoingVoluntary: Firm initiated
Z-2407-2023Class IIOngoingVoluntary: Firm initiated
Z-2408-2023Class IIOngoingVoluntary: Firm initiated
Z-2409-2023Class IIOngoingVoluntary: Firm initiated
Z-2410-2023Class IIOngoingVoluntary: Firm initiated