EMBLEM MRI S-ICD Pulse Generator Model A219; Subcutaneous Implantable Cardioverter Defibrillator
Boston Scientific Corporation began a recall of EMBLEM MRI S-ICD Pulse Generator Model A219; Subcutaneous Implantable Cardioverter Defibrillator on . FDA classified it as Class II and gives the reason as "There is a potential for a rare interaction between the EMBLEM S-ICD and LATITUDE communicator, which may cause S-ICD sensing disablement for a 24-hour interval". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-2442-2023
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Boston Scientific Corporation
- Product
- EMBLEM MRI S-ICD Pulse Generator Model A219; Subcutaneous Implantable Cardioverter Defibrillator
- Product code
- LWS Implantable Cardioverter Defibrillator (Non-Crt)
- Reason for recall
- There is a potential for a rare interaction between the EMBLEM S-ICD and LATITUDE communicator, which may cause S-ICD sensing disablement for a 24-hour interval.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Boston Scientific disseminated an Important Medical Device Information notice to the physician or healthcare professional consignees beginning on 07/11/2023 by hand delivery using their sales representatives, The notices explain the problem, risk, and recommend the following: Please distribute this letter to all HCPs who manage follow-ups of S-ICD patients. Perform the next scheduled device check in-person using a Model 3300 LATITUDE programmer with Model 3877 v1.04 software or Model 3200 EMBLEM programmer with Model 2877 v4.10 software to address any potential for this rare S-ICD behavior. Up-to-date product performance information and a device lookup tool are available within our Product Performance Resource Center at www.bostonscientific.com/ppr. If you have additional questions regarding this information or need to report an adverse event, please contact your Boston Scientific representative or Technical Services, and your local regulatory authority, as applicable. On 08/16/2023 Boston Scientific mailed the letter to EMBLEM S-ICD implanting and/or following physicians/clinics of record. On 09/15/2023, Boston Scientific began distributing a letter dated September 2023 (with the original communication enclosed), to new implanting and following physicians/clinics of record since the August 2023 communication.
- Quantity in commerce
- 41,411 units
- Distribution
- Worldwide
- Product codes and lots
- GTIN: 00802526581519, 00802526584404, 00802526584411, 00802526590405, 00802526590429, 00802526590436; All EMBLEM S-ICDs enrolled in LATITUDE
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2441-2023 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2442-2023 | Class II | Ongoing | Voluntary: Firm initiated |