Z-2489-2023Class II

Upper roller bracket (part number 1088370) for the Gen2 gantry enclosure for the Radixact Treatment Delivery...

Accuray Incorporated / initiated 2023-08-03 / Completed / root cause: design

Accuray Incorporated began a recall of Upper roller bracket (part number 1088370) for the Gen2 gantry enclosure for the Radixact Treatment Delivery System on . FDA classified it as Class II and gives the reason as "It is possible the upper roller bracket in the Radixact System gantry enclosure may not provide enough clearance for service personnel to use a jib crane for moving components while servicing the system, resulting in...". As of , the recall status on record is Completed, and MDM tags the root cause as design.

Recall record

Recall number
Z-2489-2023
Classification
Class II
Status
Completed
Initiated
Posted
Recalling firm
Accuray Incorporated
Product
Upper roller bracket (part number 1088370) for the Gen2 gantry enclosure for the Radixact Treatment Delivery System.
Product code
IYE Accelerator, Linear, Medical
Reason for recall
It is possible the upper roller bracket in the Radixact System gantry enclosure may not provide enough clearance for service personnel to use a jib crane for moving components while servicing the system, resulting in possible injury to the service personnel.
Root cause tag
design
FDA root cause
Device Design
Action
The recalling firm began issuing letters dated 7/19/2023 via hand-delivery by the field service engineer. The letter explained the issue and explained that if their Radixact System is being serviced by a third-party provider, please ensure they are also aware and take relevant precautions while servicing the system. The product impacted by the issue was detailed. The letter explained the issue arises only during the servicing of the Radixact System and customers can continue the clinical use of their system. The letter informs the consignee that Accuray will inspect the upper roller bracket of each impacted Radixact system to confirm there is adequate clearance for a jib crane. If there is sufficient clearance, then no action is necessary. If there is insufficient clearance, then the old bracket will be replaced with the new bracket following the released updated service procedures. An acknowledgment form was enclosed documenting the consignee confirms understanding the content of the field safety notice and has distributed the information to all applicable members of their staff.
Quantity in commerce
82
Distribution
Worldwide - US Nationwide distribution in the states of CA, GA, IL, LA, NJ, OR, PA, SC, SD, TX, and WA. There was government distribution and no military distribution. The countries of Australia, Belgium, Canada, Colombia, France, India, Indonesia, Italy, Japan, Kazakhstan, Malaysia, Poland, Saudi Arabia, South Korea, Spain, Switzerland, Taiwan, Thailand, Turkey, and United Kingdom.
Product codes and lots
UDI-DI number 00811376030009. Serial numbers 4010505, 4010506, 4010507, 4010508, 4010509, 4010510, 4010511, 4010513, 4010515, 4010516, 4010517, 4010518, 4010519, 4010521, 4010522, 4010523, 4010524, 4010525, 4010526, 4010527, 4010528, 4010529, 4010530, 4010531, 4010532, 4010533, 4010536, 4010537, 4010538, 4010539, 4010541, 4010542, 4010543, 4010544, 4010545, 4010546, 4010548, 4010549, 4010550, 4010551, 4010552, 4010553, 4010554, 4010555, 4010556, 4010557, 4010558, 4010559, 4010560, 4010561, 4010562, 4010563, 4010564, 4010565, 4010566, 4010567, 4010568, 4010569, 4010570, 4010571, 4010573, 4010574, 4010577, 4010579, 4010582, 4010583, 4010584, 4010585, 4010586, 4010587, 4010588, 4010589, 4010590, 4010592, 4010593, 4010594, 4010596, 4010599, 4010607, 4010600, 4010601, and 4010609.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2489-2023
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2489-2023Class IICompletedVoluntary: Firm initiated