Z-2485-2023Class II

Wireless Foot Switch distributed with Philips Allura Xper and Azurion Interventional Fluoroscopic X-Ray...

Philips Medical Systems Nederland B.V. / initiated 2023-06-30 / Open, Classified

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. began a recall of Wireless Foot Switch distributed with Philips Allura Xper and Azurion Interventional Fluoroscopic X-Ray Systems on . FDA classified it as Class II and gives the reason as "Loss of availability of the wireless foot switch during procedures". As of , the recall status on record is Open, Classified.

Recall record

Recall number
Z-2485-2023
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Philips Medical Systems Nederland B.V.
Product
Wireless Foot Switch distributed with Philips Allura Xper and Azurion Interventional Fluoroscopic X-Ray Systems
Product code
OWB Interventional Fluoroscopic X-Ray System
Reason for recall
Loss of availability of the wireless foot switch during procedures.
FDA root cause
Under Investigation by firm
Action
Customers were sent an URGENT Medical Device Correction letter dated 6/23/2023 via USPS Certified Mail. Customers are instructed to ensure that the wired footswitch is always connected to the system in case the wireless foot switch becomes unavailable. Customers are instructed to download a copy of the addendum to the IFU for the wireless foot switch and the Quick Reference Card at Http://philips.com/doc_library, and distribute these materials to those who use the system and retain a copy with the system's IFU. These documents are to be placed in an area where they are likely to be seen/viewed with system documentation. Customers are to return the attached response form to Philips no later than 30 days from receipt of the notification to [email protected]. Philips will contact customers to schedule a Field Service Engineer visit to confirm that the wired foot switch is connected to affected systems, perform a check of the charger, and provide customers with a hardcopy of the addendum. Customers with questions can contact Philips Technical Support Line at 1-800-722-9377.
Quantity in commerce
1865 units
Distribution
Worldwide distribution - US Nationwide and the countries of Australia, Austria, Bahrain, Belgium, Bosnia and Herzegovina, Brazil, Chile, China, Croatia, Czech Republic, Denmark, Egypt, Fiji, Finland, France, Germany, Greece, Hong Kong, Iceland, India, Iran, Ireland, Israel, Italy, Japan, Jordan, Korea, Republic of, Kosovo, Latvia, Lebanon, Libya, Malaysia, Maldives, Malta, Martinique, Mexico, Morocco, Netherlands, New Zealand, Norway, Oman, Palestine, State of, Poland, Portugal, Qatar, R¿union, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Tanzania, Thailand, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uzbekistan, and Yemen.
Product codes and lots
Model No.: 722001, 722003, 722005, 722006, 722008, 722010, 722011, 722012, 722013, 722015, 722023, 722025, 722026, 722027, 722028, 722029, 722033, 722035, 722038, 722039, 722058, 722059, 722064, 722065, 722066, 722067, 722068, 722078, 722079; No UDI-DI; Lot No. 459800772231, 459800772233, 459800772261, 459800772263, 459800415531,459800415532, 459800415533, 459800415534, 459800415535, 459800415571, 459800415572, 459800415573, 459800415574, 459800415575.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2485-2023
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2485-2023Class IIOngoingVoluntary: Firm initiated