Z-2472-2023Class II

Atellica CH LDL Cholesterol- In vitro diagnostic use in the quantitative determination of LDL cholesterol in...

Siemens Healthcare Diagnostics Inc. / initiated 2023-07-10 / Open, Classified / root cause: manufacturing process

Siemens Healthcare Diagnostics, Inc. began a recall of Atellica CH LDL Cholesterol- In vitro diagnostic use in the quantitative determination of LDL cholesterol in human serum and plasma (lithium heparin, EDTA... on . FDA classified it as Class II and gives the reason as "Potential for reagent carryover on the Atellica CH 930 resulting in a positive bias that could impact quality control (QC), patient samples, and calibrator results with from Atellica CH Total Bilirubin_2 (TBil_2)...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2472-2023
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
Atellica CH LDL Cholesterol- In vitro diagnostic use in the quantitative determination of LDL cholesterol in human serum and plasma (lithium heparin, EDTA, sodium heparin Test Code: LDLC Siemens Material Number (SMN): 11537214
Product code
JFM Enzymatic Method, Bilirubin
Reason for recall
Potential for reagent carryover on the Atellica CH 930 resulting in a positive bias that could impact quality control (QC), patient samples, and calibrator results with from Atellica CH Total Bilirubin_2 (TBil_2), Atellica CH LDL Cholesterol (LDLC), Atellica CH Gamma-Glutamyl Transferase (GGT) and Atellica CH HDL Cholesterol (HDLC)
Root cause tag
manufacturing process
FDA root cause
Under Investigation by firm
Action
Siemens Healthineers issued Urgent Medical Device Correction ACHC23-01.C.US to US affected customers via FedEx beginning on 07/10/2023; and Urgent Field Safety Notice ACHC23-01.C.OUS,OUS Siemens reporting sites with affected customers, for distribution and regional reporting to their local competent authorities beginning on 07/10/2023. Letter states reason for recall, health risk and action to take: Please review this letter with your Medical Director to determine the appropriate course of action, including for any previously generated results, if applicable. Perform the instructions provided in the Additional Information section. Complete and return the Field Correction Effectiveness Check Form attached to this letter within 30 days. If you have received any complaints of illness or adverse events associated with the products listed in Table 1, immediately contact your local Siemens Healthineers Customer Care Center or your local Siemens Healthineers technical support representative. Please retain this letter with your laboratory records and forward this letter to those who may have received this product. If you have any questions, please contact your Siemens Healthineers Customer Care Center or your local Siemens Healthineers technical support representative. If your laboratory has multiple Atellica CH 930 Analyzers, separate the assays as follows: " Perform testing of TBil_2 on a separate analyzer from Glucose Oxidase (GluO) " Perform testing of LDLC on a separate analyzer from Lipase (Lip) " Perform testing of GGT on a separate analyzer from Magnesium (Mg) " Perform testing of HDLC on a separate analyzer from Uric Acid (UA) If you choose not to or are unable to separate the assays as indicated above, batch testing of GluO, Lip, Mg and UA may be performed. Note: When batch testing impacted assays, GluO, Lip, Mg and UA, an RPC2 wash mitigation must be initiated after completion of tests with the assay causing carryover, TBil_2, LDLC, GGT and HDLC, respectivel
Quantity in commerce
6733 units
Distribution
US Nationwide - Worldwide Distribution Foreign: Taiwan Uruguay Vietnam
Product codes and lots
Unique Device Identification (UDI): 00630414611037 All Lot Numbers

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2472-2023
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2471-2023Class IIOngoingVoluntary: Firm initiated
Z-2472-2023Class IIOngoingVoluntary: Firm initiated
Z-2473-2023Class IIOngoingVoluntary: Firm initiated
Z-2474-2023Class IIOngoingVoluntary: Firm initiated