Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2479-2023Class II

PASCAL Precision Transcatheter Valve Repair System, Models: 20000ISM and 20000IS

Edwards Lifesciences LLC / initiated 2023-07-27 / Open, Classified / root cause: manufacturing process

Edwards Lifesciences, LLC began a recall of PASCAL Precision Transcatheter Valve Repair System, Models: 20000ISM and 20000IS on . FDA classified it as Class II and gives the reason as "Due to a manufacturing issue, when using valve repair systems, intended to repair insufficient mitral and/or tricuspid valves, users experienced resistance when attempting to release the implant from the delivery system...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2479-2023
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Edwards Lifesciences LLC
Product
PASCAL Precision Transcatheter Valve Repair System, Models: 20000ISM and 20000IS
Product code
NKM Mitral Valve Repair Devices
Reason for recall
Due to a manufacturing issue, when using valve repair systems, intended to repair insufficient mitral and/or tricuspid valves, users experienced resistance when attempting to release the implant from the delivery system by rotating the release knob. Additional torque required to initiate rotation of the release wire from the implant, could cause implant rotation or additional implant implantation.
Root cause tag
manufacturing process
FDA root cause
Process change control
Action
On 7/27/23, recall notices were sent to customers and distributors who were asked to do the following: 1) Stop using and work with your firm representative to return affected devices. 2) Share this notice with the appropriate clinical staff at your site. 3) Complete and return the customer acknowledgement form. 4) Distributors: Notify your customers by sending this customer notification to any of your customers who have purchased the impacted devices.
Quantity in commerce
92
Distribution
International distribution to the countries of Austria, Belgium, Switzerland, Germany, Cyprus, Israel, UK, Greece, Ireland, Italy, Luxembourg, Poland.
Product codes and lots
Basic-UDI Code: 0690103S004PAS000BC. Model/Serial Numbers: 20000ISM/10844549, 10844558, 10921747, 10700424, 10809849, 10809852, 10844564, 10864947, 10903936, 10903937, 10921752, 10921754, 10865580, 10921757, 10921758, 10921760, 10700439, 10776944, 10776966, 10776982, 10809844, 10844243, 10844251, 10844281, 10844292, 10844571, 10848756, 10864921, 10864926, 10864927, 10864942, 10864943, 10864945, 10864950, 10865585, 10865595, 10865603, 10865652, 10865653, 10865654, 10865656, 10865720, 10865723, 10865724, 10865730, 10865731, 10903944, 10903960, 10921729, 10921739, 10921922, 10921924, 10921925, 10921932, 10921941, 10844264, 10844284, 10844288, 10844581, 10844585, 10848729, 10848730, 10848739, 10848740, 10848744, 10848746, 10848747, 10848750, 10848753, 10864833, 10864838, 10864840, 10864844, 10864897, 10865584, 10865613, 10865616, 10865618, 10865649, 10865719, 10865722, 10865727, 10921733, 10921734, 10921918, 10951672, 10951679, 10951683, 10951685, 10951687, 10951691, 10865732, 10865735, 10903975, 10809836, 10848723, 10921930, 10844300, 10844302, 10844305, 10864959, 10864960, 10921926, 10921927, 10844551, 10844563, 10844566, 10853110, 10903977, 10921748, 10921750, 10922081, 10922089, 10922093, 10922096, 10922098, 10922102, 10922103, 10922106, 10922107, 10922108, 10922110, 10922111, 10922116, 10922118, 10922119, 10922121, 10922122, 10922126, 10922127, 10922129, 10922133, 10922138, 10922140, 10922142, 10951734, 10951737, 10951740, 10951741, 10951743, 10951744, 10951745, 10951747, 10951748, 10951749, 10951750, 10951751, 10951753, 10951754, 10951756, 10951757, 10951758, 10951760, 10951761, 10951763, 10969562, 10969572, 10969573, 10969576, 10969577, 10969579, 10969583, 10969587, 10969591, 10969592, 10969593, 10969595, 10969597, 10969598, 10969599, 10969601, 10969603, 10969606, 10969611, 10969616, 10969619, 10969620, 10969625, 10969629, 10848722, 10865736, 10904022, 10904024, 10904026, 10904028, 10904029, 10864924, 10864928, 10864931, 10864936, 10864939, 10865549, 10865716, 10921861. 20000IS/10875150, 10853171, 10853233, 10853196, 10853203, 10853208, 10853213, 10853222, 10853235, 10875121, 10875124, 10875126, 10875131, 10875137, 10875138, 10875143, 10875145, 10875146, 10875152, 10853206, 10875154, 10875123, 10875136, 10875142, 10875144

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2479-2023
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2479-2023Class IIOngoingVoluntary: Firm initiated