PASCAL Precision Transcatheter Valve Repair System, Models: 20000ISM and 20000IS
Edwards Lifesciences, LLC began a recall of PASCAL Precision Transcatheter Valve Repair System, Models: 20000ISM and 20000IS on . FDA classified it as Class II and gives the reason as "Due to a manufacturing issue, when using valve repair systems, intended to repair insufficient mitral and/or tricuspid valves, users experienced resistance when attempting to release the implant from the delivery system...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2479-2023
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Edwards Lifesciences LLC
- Product
- PASCAL Precision Transcatheter Valve Repair System, Models: 20000ISM and 20000IS
- Product code
- NKM Mitral Valve Repair Devices
- Reason for recall
- Due to a manufacturing issue, when using valve repair systems, intended to repair insufficient mitral and/or tricuspid valves, users experienced resistance when attempting to release the implant from the delivery system by rotating the release knob. Additional torque required to initiate rotation of the release wire from the implant, could cause implant rotation or additional implant implantation.
- Root cause tag
- manufacturing process
- FDA root cause
- Process change control
- Action
- On 7/27/23, recall notices were sent to customers and distributors who were asked to do the following: 1) Stop using and work with your firm representative to return affected devices. 2) Share this notice with the appropriate clinical staff at your site. 3) Complete and return the customer acknowledgement form. 4) Distributors: Notify your customers by sending this customer notification to any of your customers who have purchased the impacted devices.
- Quantity in commerce
- 92
- Distribution
- International distribution to the countries of Austria, Belgium, Switzerland, Germany, Cyprus, Israel, UK, Greece, Ireland, Italy, Luxembourg, Poland.
- Product codes and lots
- Basic-UDI Code: 0690103S004PAS000BC. Model/Serial Numbers: 20000ISM/10844549, 10844558, 10921747, 10700424, 10809849, 10809852, 10844564, 10864947, 10903936, 10903937, 10921752, 10921754, 10865580, 10921757, 10921758, 10921760, 10700439, 10776944, 10776966, 10776982, 10809844, 10844243, 10844251, 10844281, 10844292, 10844571, 10848756, 10864921, 10864926, 10864927, 10864942, 10864943, 10864945, 10864950, 10865585, 10865595, 10865603, 10865652, 10865653, 10865654, 10865656, 10865720, 10865723, 10865724, 10865730, 10865731, 10903944, 10903960, 10921729, 10921739, 10921922, 10921924, 10921925, 10921932, 10921941, 10844264, 10844284, 10844288, 10844581, 10844585, 10848729, 10848730, 10848739, 10848740, 10848744, 10848746, 10848747, 10848750, 10848753, 10864833, 10864838, 10864840, 10864844, 10864897, 10865584, 10865613, 10865616, 10865618, 10865649, 10865719, 10865722, 10865727, 10921733, 10921734, 10921918, 10951672, 10951679, 10951683, 10951685, 10951687, 10951691, 10865732, 10865735, 10903975, 10809836, 10848723, 10921930, 10844300, 10844302, 10844305, 10864959, 10864960, 10921926, 10921927, 10844551, 10844563, 10844566, 10853110, 10903977, 10921748, 10921750, 10922081, 10922089, 10922093, 10922096, 10922098, 10922102, 10922103, 10922106, 10922107, 10922108, 10922110, 10922111, 10922116, 10922118, 10922119, 10922121, 10922122, 10922126, 10922127, 10922129, 10922133, 10922138, 10922140, 10922142, 10951734, 10951737, 10951740, 10951741, 10951743, 10951744, 10951745, 10951747, 10951748, 10951749, 10951750, 10951751, 10951753, 10951754, 10951756, 10951757, 10951758, 10951760, 10951761, 10951763, 10969562, 10969572, 10969573, 10969576, 10969577, 10969579, 10969583, 10969587, 10969591, 10969592, 10969593, 10969595, 10969597, 10969598, 10969599, 10969601, 10969603, 10969606, 10969611, 10969616, 10969619, 10969620, 10969625, 10969629, 10848722, 10865736, 10904022, 10904024, 10904026, 10904028, 10904029, 10864924, 10864928, 10864931, 10864936, 10864939, 10865549, 10865716, 10921861. 20000IS/10875150, 10853171, 10853233, 10853196, 10853203, 10853208, 10853213, 10853222, 10853235, 10875121, 10875124, 10875126, 10875131, 10875137, 10875138, 10875143, 10875145, 10875146, 10875152, 10853206, 10875154, 10875123, 10875136, 10875142, 10875144
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2479-2023 | Class II | Ongoing | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.