Reusable distal attachments Models: MH-462, MH-463, MH-464, MH-465, MH-466, MH-483, MH-587, MH-588(US)...
Aomori Olympus Co., Ltd. began a recall of Reusable distal attachments Models: MH-462, MH-463, MH-464, MH-465, MH-466, MH-483, MH-587, MH-588(US), MH-589, MH-590, MH-591, MH-592, MH-593, MH-594, MH-595... on . FDA classified it as Class II and gives the reason as "The Distal Attachment contain di-2-ethylhexyl phthalate (DEHP), which exceeds the permissible exposure level calculated according to ISO 10993-17: 2002". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2523-2023
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- AOMORI OLYMPUS CO., Ltd. Kuroishi Site
- Product
- Reusable distal attachments Models: MH-462, MH-463, MH-464, MH-465, MH-466, MH-483, MH-587, MH-588(US), MH-589, MH-590, MH-591, MH-592, MH-593, MH-594, MH-595, MH-596, MH-597, MH-598, MAJ-289, MAJ-290, MAJ-291, MAJ-292, MAJ-293, MAJ-294, MAJ-295, MAJ-296, MAJ-297, MAJ-663, MAJ-664, MAJ-665, MAJ-666
- Product code
- FDS Gastroscope And Accessories, Flexible/Rigid
- Reason for recall
- The Distal Attachment contain di-2-ethylhexyl phthalate (DEHP), which exceeds the permissible exposure level calculated according to ISO 10993-17: 2002
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Olympus issued Urgent Medical Device Recall Letter dated July 27, 2023 to ENDOSCOPY DEPARTMENT, RISK MANAGEMENT. Letter states reason for recall, health risk and action to take: Immediately assess any product you have to identify the product number with affected lot listed in this communication, cease use and quarantine any affected products. The image below depicts the area where the lot number is identified. The lot# is on the Carton box and Pack. Please refer to Attachment 2 for the identification of the subject products. 2. Call your Olympus customer service representative at 1-800-848-9024, option 2. Olympus will issue a Return Material Authorization to return any affected product at no charge to you. Olympus will issue a credit to your facility for your affected product. 3. Olympus requests that you acknowledge receipt of this letter through the Olympus recall portal. a. Go to https://olympusamerica.com/recall b. Enter the file (recall) number: 0429 4. If you have further distributed this product, identify your customers, forward them this notification, and appropriately document your notification process. Olympus requests that you report complaints, including patient physical condition suspected to be affected by DEHP, to Olympus. Please report complaints to our Technical Assistance Center (TAC) at 1-800-848-9024 (option 1). If you require additional information, please contact Tara Safi, Field Corrective Actions Administrator at [email protected].
- Quantity in commerce
- 1,252 units
- Distribution
- US Nationwide distribution.
- Product codes and lots
- Model: UDI: MH-462 4953170047732 MH-463 4953170047756 MH-464 4953170047770 MH-465 4953170047794 MH-466 4953170047817 MH-483 N/A MH-587 4953170047855 MH-588 4953170364518 (US) MH-589 4953170047893 MH-590 N/A MH-591 4953170047930 MH-592 N/A MH-593 4953170047978 MH-594 4953170047992 MH-595 4953170048012 MH-596 N/A MH-597 4953170048050 MH-598 N/A MAJ-289 N/A MAJ-290 N/A MAJ-291 N/A MAJ-292 N/A MAJ-293 N/A MAJ-294 N/A MAJ-295 N/A MAJ-296 N/A MAJ-297 N/A MAJ-663 N/A MAJ-664 N/A MAJ-665 N/A MAJ-666 N/A DIstal Attachment- All lots
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2522-2023 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2523-2023 | Class II | Ongoing | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.
