Link OptiStem Rasp Handle-Used with the OptiStem Rasp to prepare the femoral medullary canal for a knee...

Waldemar Link GmbH & Co. KG / initiated 2023-07-26 / Open, Classified / root cause: design

Waldemar Link GmbH & Co. KG (Mfg Site) began a recall of Link OptiStem Rasp Handle-Used with the OptiStem Rasp to prepare the femoral medullary canal for a knee prosthesis implant during a total knee arthroplasty... on . FDA classified it as Class II and gives the reason as "A crack in the weld of the rasp handle may lead to a fracture of the frontal sleeve when the rasp is further impacted or extracted. The rasp remains intact. The rasp would need to be pulled out by a clamp leading to a...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.

Recall record

Recall number
Z-2481-2023
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Waldemar Link GmbH & Co. KG
Product
Link OptiStem Rasp Handle-Used with the OptiStem Rasp to prepare the femoral medullary canal for a knee prosthesis implant during a total knee arthroplasty surgical procedure Item Number: 17-5220/01
Product code
LXH Orthopedic Manual Surgical Instrument
Reason for recall
A crack in the weld of the rasp handle may lead to a fracture of the frontal sleeve when the rasp is further impacted or extracted. The rasp remains intact. The rasp would need to be pulled out by a clamp leading to a prolongation of surgery or modified surgery
Root cause tag
design
FDA root cause
Device Design
Action
LinkBio Corp.(US Importer) issued Urgent: Medical Device Recall letter via email on July 27, 203 to US Distributors . Letter states reason for recall, health risk and action to take: Please respond using the attached Distributor Reply form to indicate your understanding of this Device Recall and of the need to return the affected product. Pre-paid shipping labels will be provided once the Distributor Reply Form is returned. Please return the product as soon as possible to the following address: LinkBio Corporation 69 King St Dover, NJ 07801 Please contact LinkBio at 973-625-1333 or [email protected] for any questions.
Quantity in commerce
6 units
Distribution
IN,GA, TX
Product codes and lots
UDI-DI: 04026575284658 Lot Codes: C010108, C151519

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2481-2023
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2481-2023Class IIOngoingVoluntary: Firm initiated