Z-2590-2023Class II
Arrow ErgoPack Pressure Injectable One-Lumen PICC, REF CDC-05541-HPK1A; catheter, intravascular
Arrow International LLC (subsidiary of Teleflex Incorporated) / initiated 2023-08-10 / Open, Classified
ARROW INTERNATIONAL Inc. began a recall of Arrow ErgoPack Pressure Injectable One-Lumen PICC, REF CDC-05541-HPK1A; catheter, intravascular on . FDA classified it as Class II and gives the reason as "This recall has been initiated due to reports that the product code and product name were incorrectly listed on the lidstock of affected kits". As of , the recall status on record is Open, Classified.
Recall record
- Recall number
- Z-2590-2023
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Arrow International LLC (subsidiary of Teleflex Incorporated)
- Product
- Arrow ErgoPack Pressure Injectable One-Lumen PICC, REF CDC-05541-HPK1A; catheter, intravascular
- Product code
- LJS Catheter, Intravascular, Therapeutic, Long-Term Greater Than 30 Days
- Reason for recall
- This recall has been initiated due to reports that the product code and product name were incorrectly listed on the lidstock of affected kits.
- FDA root cause
- Under Investigation by firm
- Action
- Teleflex/Arrow International issue an URGENT MEDICAL DEVICE RECALL notice on 08/10/2023 by FedEx 2nd day mail. The letter explained the problem, risk to patient, and requested the following actions: Medical facilities: Immediately cease use and distribution of affected product and quarantine for return. Distributors: Immediately cease use and distribution of affected product and quarantine for return. Provide a copy of this recall notice to all customers who have received impacted product. Request they complete the Acknowledgement Form and return it to you. Firm is requesting return of the affected product.
- Quantity in commerce
- 190 units
- Distribution
- US Nationwide distribution.
- Product codes and lots
- Batch/Lot number 13F22J0617
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2590-2023 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2591-2023 | Class II | Ongoing | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.