Z-2404-2023Class II

MultiDiagnost-Eleva

Philips Medical Systems Nederland B.V. / initiated 2023-07-19 / Open, Classified

Philips Medical Systems Nederland B.V. began a recall of MultiDiagnost-Eleva on . FDA classified it as Class II and gives the reason as "A foot switch pedal may get stuck in the active position when the user releases the pedal, resulting in the emission of unintended radiation". As of , the recall status on record is Open, Classified.

Recall record

Recall number
Z-2404-2023
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Philips Medical Systems Nederland B.V.
Product
MultiDiagnost-Eleva
Product code
KPR System, X-Ray, Stationary
Reason for recall
A foot switch pedal may get stuck in the active position when the user releases the pedal, resulting in the emission of unintended radiation
FDA root cause
Radiation Control for Health and Safety Act
Action
Philips will send a customer notification to the existing users of the affected systems to alert them of the issue and provide actions that should be taken by the customer/user in order to prevent risks to patients. Philips will contact customers to schedule a visit to inspect the foot switch and provide an addendum to the Instructions for Use (IFU) of the system detailing the appropriate cleaning and maintenance instructions. (Philips reference 2023-IGT-BST-011).
Quantity in commerce
5432 systems in total
Distribution
US Nationwide Distribution
Product codes and lots
MultiDiagnost-Eleva

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2404-2023
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2402-2023Class IIOngoingFDA Mandated
Z-2403-2023Class IIOngoingFDA Mandated
Z-2404-2023Class IIOngoingFDA Mandated