Z-0153-2024Class II

Preat REF 9007123, Neodent GM x 4mm Healing Abutment, Rx Only, Non-Sterile, Single Use Only. Dental

PREAT Corporation / initiated 2023-07-18 / Open, Classified

Preat Corp began a recall of Preat REF 9007123, Neodent GM x 4mm Healing Abutment, Rx Only, Non-Sterile, Single Use Only. Dental on . FDA classified it as Class II and gives the reason as "The healing abutments contain a dimensional condition that may lead to a gap between the healing abutment and the implant allowing for microleakage". As of , the recall status on record is Open, Classified.

Recall record

Recall number
Z-0153-2024
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
PREAT Corporation
Product
Preat REF 9007123, Neodent GM x 4mm Healing Abutment, Rx Only, Non-Sterile, Single Use Only. Dental
Product code
NHA Abutment, Implant, Dental, Endosseous
Reason for recall
The healing abutments contain a dimensional condition that may lead to a gap between the healing abutment and the implant allowing for microleakage
FDA root cause
Process change control
Action
On August 4, 2023, Preat Corporation issued a "Urgent Medical Device Recall" notification via: USPS. Preat ask consignees to take the following actions: Product within the scope of the recall shall be discarded or returned to Preat Corporation. Instructions for the return will be provided. Complete and email response form to Nichole Fetterman, E-Mail: [email protected] or mail to Preat Corporation, Attn. Nichole Fetterman, 2625 Skyway Drive, Ste. B, Santa Maria, CA 93455
Quantity in commerce
23 units
Distribution
US: FL, CA, OR, TX, OH, CA, SC, OR, NJ, SC, OUS: None
Product codes and lots
UDI-DI: 00842092161388/Lot #OF018357

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0153-2024
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0152-2024Class IIOngoingVoluntary: Firm initiated
Z-0153-2024Class IIOngoingVoluntary: Firm initiated
Z-0154-2024Class IIOngoingVoluntary: Firm initiated