Hamilton-T1, REF: MSP161339/02, ESM Shielding Set, Spare Part for ventilator
Hamilton Medical, Inc. began a recall of Hamilton-T1, REF: MSP161339/02, ESM Shielding Set, Spare Part for ventilator on . FDA classified it as Class I and gives the reason as "Degrading capacitors on the control board of ventilators and spare parts may leak electrolyte onto the control board causing a short circuit on the board and/or the capacitor to lose function and lead to interruption of...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2530-2023
- Classification
- Class I
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Hamilton Medical AG
- Product
- Hamilton-T1, REF: MSP161339/02, ESM Shielding Set, Spare Part for ventilator
- Product code
- CBK Ventilator, Continuous, Facility Use
- Reason for recall
- Degrading capacitors on the control board of ventilators and spare parts may leak electrolyte onto the control board causing a short circuit on the board and/or the capacitor to lose function and lead to interruption of ventilation.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- On July 26, 2023, Hamilton Medical issued Urgent Medical Device Correction via E-Mail. Hamilton Medical asked consignees to take the following actions 1. Check for affected devices in your facility. Get in contact with your local distribution partner and get the devices serviced with high priority. 2. Please sign the enclosed/attached sheet to confirm that, as an End- Customer, you have received and duly noted this Medical Device Correction.
- Quantity in commerce
- 4 shielding set
- Distribution
- Distribution US nationwide, including Puerto Rico. OUS: Not provided
- Product codes and lots
- UDI: 07630002800761 / Affected Serial Numbers: 4050193, 4050210, 4050194 and 4016669
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2525-2023 | Class I | Ongoing | Voluntary: Firm initiated | ||
| Z-2526-2023 | Class I | Ongoing | Voluntary: Firm initiated | ||
| Z-2527-2023 | Class I | Ongoing | Voluntary: Firm initiated | ||
| Z-2528-2023 | Class I | Ongoing | Voluntary: Firm initiated | ||
| Z-2529-2023 | Class I | Ongoing | Voluntary: Firm initiated | ||
| Z-2530-2023 | Class I | Ongoing | Voluntary: Firm initiated |