Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2659-2023Class II

Veran:SPiN Thoracic Navigation System Models: SYS-4230 SPiN Thoracic Navigation System - 230V 50Hz SYS-4000...

OLYMPUS Corporation of the Americas / initiated 2023-08-10 / Open, Classified / root cause: manufacturing process

Olympus Corporation of the Americas began a recall of Veran:SPiN Thoracic Navigation System Models: SYS-4230 SPiN Thoracic Navigation System - 230V 50Hz SYS-4000 SPiN Thoracic Navigation System SYS-3230 ig4... on . FDA classified it as Class II and gives the reason as "Non-conformances in Veran devices (ig4 Image Guided System, SPiN Thoracic Navigation System, SPiN Planning Laptop Workstation, and all associated navigation instruments) manufacturing to design requirements and...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2659-2023
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
OLYMPUS Corporation of the Americas
Product
Veran:SPiN Thoracic Navigation System Models: SYS-4230 SPiN Thoracic Navigation System - 230V 50Hz SYS-4000 SPiN Thoracic Navigation System SYS-3230 ig4 System, 230V, 50Hz / SPiN Thoracic Navigation System 0095 SYS-3000 SPiN Thoracic Navigation System SYS-2400 SPiN Drive System SYS-2230 ig4 System V2, 230V, 50Hz SYS-2000 ig4 Image Guides System SYS-1500 SPiN View" System SYS-1000 ig4 Image Guides System SYS-0220 ig4 System, 220V, 50Hz SYS-0185 SPiN Planning Laptop Workstation SYS-0128 Planning Workstation w/Monitor SYS-0002 ig4 System INS-7150 SPiN Vision Scope Adapter
Product code
JAK System, X-Ray, Tomography, Computed
Reason for recall
Non-conformances in Veran devices (ig4 Image Guided System, SPiN Thoracic Navigation System, SPiN Planning Laptop Workstation, and all associated navigation instruments) manufacturing to design requirements and insufficient evidence to demonstrate adequate electromagnetic compatibility (EMC). EMC is the ability of electrical equipment and systems to function acceptably without interference and to not interfere with other devices within its operating environment
Root cause tag
manufacturing process
FDA root cause
Under Investigation by firm
Action
Olympus Corporation of the Americas ( Olympus ),[Veran Medical Technologies, a wholly owned subsidiary] issued Urgent Medical Device Recall Letter on 8/10/23 to: Endoscopy Department, Pulmonary Department, Risk Management for the VERAN ig4 Image Guided System, SPiN Thoracic Navigation System, SPiN Planning Laptop Workstation, and all associated navigation instruments. Letter states reason for recall, health risk and action to take: Veran requests you to take the following actions: 1. Inspect your inventory and identify any Veran models. Please check all areas of the hospital to determine if any of these devices remain in inventory. 2. Call your Olympus customer service representative at 1-800-848-9024, option 2. Olympus will issue a Return Material Authorization to return any affected product at no charge. Your Olympus Customer Service representative will advise you on reimbursement for your Veran devices. 3. Olympus requests that you acknowledge receipt of this letter through the Olympus recall portal. a. Go to https://olympusamerica.com/recall b. Enter the file (recall) number: 0430 4. If you have further distributed this product, identify your customers, forward them this notification, and appropriately document your notification process. If you require additional information, please contact Tara Safi, Field Corrective Actions Administrator at [email protected].
Quantity in commerce
226 units
Distribution
Worldwide distribution - US Nationwide and the countries of CN, DE, HK, IT, KR, SG, TH, TW.
Product codes and lots
Model-UDI: SYS-4230 SPiN Thoracic Navigation System - 230V 50Hz 00815686020118; SYS-4000 SPiN Thoracic Navigation System 00815686020101; SYS-3230 ig4 System, 230V, 50Hz / SPiN Thoracic Navigation System 00815686020095; SYS-3000 SPiN Thoracic Navigation System 00815686020088; SYS-2400 SPiN Drive System N/A; SYS-2230 ig4 System V2, 230V, 50Hz 00815686020071; SYS-2000 ig4 Image Guides System 00815686020064; SYS-1500 SPiN View" System 00815686020057; SYS-1000 ig4 Image Guides System N/A; SYS-0220 ig4 System, 220V, 50Hz N/A; SYS-0185 SPiN Planning Laptop Workstation 00815686020033; SYS-0128 Planning Workstation w/Monitor N/A; SYS-0002 ig4 System N/A; INS-7150 SPiN Vision Scope Adapter N/A; All serial numbers and all lot numbers.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2659-2023
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2659-2023Class IIOngoingVoluntary: Firm initiated
Z-2660-2023Class IIOngoingVoluntary: Firm initiated
Z-2661-2023Class IIOngoingVoluntary: Firm initiated
Z-2662-2023Class IIOngoingVoluntary: Firm initiated
Z-2663-2023Class IIOngoingVoluntary: Firm initiated
Z-2664-2023Class IIOngoingVoluntary: Firm initiated