Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2660-2023Class II

Veran Endobronchial: Models: INS-5925 SPiN Access Catheter 180 Degree for use w/ Pentax Scopes INS-5920 SPiN...

OLYMPUS Corporation of the Americas / initiated 2023-08-10 / Open, Classified / root cause: manufacturing process

Olympus Corporation of the Americas began a recall of Veran Endobronchial: Models: INS-5925 SPiN Access Catheter 180 Degree for use w/ Pentax Scopes INS-5920 SPiN Access Catheter 90 Degree for use w/ Pentax Scopes... on . FDA classified it as Class II and gives the reason as "Non-conformances in Veran devices (ig4 Image Guided System, SPiN Thoracic Navigation System, SPiN Planning Laptop Workstation, and all associated navigation instruments) manufacturing to design requirements and...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2660-2023
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
OLYMPUS Corporation of the Americas
Product
Veran Endobronchial: Models: INS-5925 SPiN Access Catheter 180 Degree for use w/ Pentax Scopes INS-5920 SPiN Access Catheter 90 Degree for use w/ Pentax Scopes INS-5915 SPiN Access Catheter 180 for use w/ Olympus Scopes INS-5910 SPiN Access Catheter 90 for use w/ Olympus 190 Scopes INS-5905 SPiN Access Catheter 180 for use w/ Olympus Scopes INS-5900 SPiN Access Catheter 90 Degree for use w/ Olympus Scopes INS-5700 SPiN EWC Biopsy Guide Kit
Product code
JAK System, X-Ray, Tomography, Computed
Reason for recall
Non-conformances in Veran devices (ig4 Image Guided System, SPiN Thoracic Navigation System, SPiN Planning Laptop Workstation, and all associated navigation instruments) manufacturing to design requirements and insufficient evidence to demonstrate adequate electromagnetic compatibility (EMC). EMC is the ability of electrical equipment and systems to function acceptably without interference and to not interfere with other devices within its operating environment
Root cause tag
manufacturing process
FDA root cause
Under Investigation by firm
Action
Olympus Corporation of the Americas ( Olympus ),[Veran Medical Technologies, a wholly owned subsidiary] issued Urgent Medical Device Recall Letter on 8/10/23 to: Endoscopy Department, Pulmonary Department, Risk Management for the VERAN ig4 Image Guided System, SPiN Thoracic Navigation System, SPiN Planning Laptop Workstation, and all associated navigation instruments. Letter states reason for recall, health risk and action to take: Veran requests you to take the following actions: 1. Inspect your inventory and identify any Veran models. Please check all areas of the hospital to determine if any of these devices remain in inventory. 2. Call your Olympus customer service representative at 1-800-848-9024, option 2. Olympus will issue a Return Material Authorization to return any affected product at no charge. Your Olympus Customer Service representative will advise you on reimbursement for your Veran devices. 3. Olympus requests that you acknowledge receipt of this letter through the Olympus recall portal. a. Go to https://olympusamerica.com/recall b. Enter the file (recall) number: 0430 4. If you have further distributed this product, identify your customers, forward them this notification, and appropriately document your notification process. If you require additional information, please contact Tara Safi, Field Corrective Actions Administrator at [email protected].
Quantity in commerce
6484 units
Distribution
Worldwide distribution - US Nationwide and the countries of CN, DE, HK, IT, KR, SG, TH, TW.
Product codes and lots
Model- UDI: INS-5925 SPiN Access Catheter¿ 180 Degree for use w/ Pentax Scopes 00815686021313; INS-5920 SPiN Access Catheter¿ 90 Degree for use w/ Pentax Scopes 00815686021306; INS-5915 SPiN Access Catheter¿ 180 for use w/ Olympus Scopes 00815686021290; INS-5910 SPiN Access Catheter¿ 90 for use w/ Olympus 190 Scopes 00815686021276; INS-5905 SPiN Access Catheter¿ 180 for use w/ Olympus Scopes 00815686021269; INS-5900 SPiN Access Catheter¿ 90 Degree for use w/ Olympus Scopes 00815686021153; INS-5700 SPiN EWC Biopsy Guide Kit 00815686021108. All serial numbers and all lot numbers

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2660-2023
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2659-2023Class IIOngoingVoluntary: Firm initiated
Z-2660-2023Class IIOngoingVoluntary: Firm initiated
Z-2661-2023Class IIOngoingVoluntary: Firm initiated
Z-2662-2023Class IIOngoingVoluntary: Firm initiated
Z-2663-2023Class IIOngoingVoluntary: Firm initiated
Z-2664-2023Class IIOngoingVoluntary: Firm initiated