Z-2663-2023Class II

Veran : Percutaneous Always on Track Models: INS-5040 Always-On vTrack Universal Tracker, LG 7-12ga INS-5039...

OLYMPUS Corporation of the Americas / initiated 2023-08-10 / Open, Classified / root cause: manufacturing process

Olympus Corporation of the Americas began a recall of Veran : Percutaneous Always on Track Models: INS-5040 Always-On vTrack Universal Tracker, LG 7-12ga INS-5039 Always-On vTrack Universal Tracker, SM/CL, 12-18ga... on . FDA classified it as Class II and gives the reason as "Non-conformances in Veran devices (ig4 Image Guided System, SPiN Thoracic Navigation System, SPiN Planning Laptop Workstation, and all associated navigation instruments) manufacturing to design requirements and...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2663-2023
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
OLYMPUS Corporation of the Americas
Product
Veran : Percutaneous Always on Track Models: INS-5040 Always-On vTrack Universal Tracker, LG 7-12ga INS-5039 Always-On vTrack Universal Tracker, SM/CL, 12-18ga INS-5036 "22 UTW 125mm Always-On Tip Tracked Biopsy Needle, Trocar point" INS-5034 17 UTW Lg 123mm Always-On Tip Tracked Biopsy Needle, Chiba point 00815686020354 INS-5032 "19 UTW 150mm Always-On Tip Tracked Biopsy Needle, Chiba point" INS-5029 "19 UTW 105 mm Always-On Tip Tracked biopsy needle, Chiba oint" INS-5028 19 UTW 155 mm Always-On Tip Tracked biopsy needle, Chiba point INS-5024 "17ga 150mm Always-On Tip Tracked Biopsy Needle, Chiba point" INS-5023 17ga 100mm Always-On Tip Tracked Biopsy Needle, Chiba point INS-5017 17ga x 155mm Tip Tracked Bx Needle INS-5016 17ga x 155mm Tip Tracked Bx Needle
Product code
JAK System, X-Ray, Tomography, Computed
Reason for recall
Non-conformances in Veran devices (ig4 Image Guided System, SPiN Thoracic Navigation System, SPiN Planning Laptop Workstation, and all associated navigation instruments) manufacturing to design requirements and insufficient evidence to demonstrate adequate electromagnetic compatibility (EMC). EMC is the ability of electrical equipment and systems to function acceptably without interference and to not interfere with other devices within its operating environment
Root cause tag
manufacturing process
FDA root cause
Under Investigation by firm
Action
Olympus Corporation of the Americas ( Olympus ),[Veran Medical Technologies, a wholly owned subsidiary] issued Urgent Medical Device Recall Letter on 8/10/23 to: Endoscopy Department, Pulmonary Department, Risk Management for the VERAN ig4 Image Guided System, SPiN Thoracic Navigation System, SPiN Planning Laptop Workstation, and all associated navigation instruments. Letter states reason for recall, health risk and action to take: Veran requests you to take the following actions: 1. Inspect your inventory and identify any Veran models. Please check all areas of the hospital to determine if any of these devices remain in inventory. 2. Call your Olympus customer service representative at 1-800-848-9024, option 2. Olympus will issue a Return Material Authorization to return any affected product at no charge. Your Olympus Customer Service representative will advise you on reimbursement for your Veran devices. 3. Olympus requests that you acknowledge receipt of this letter through the Olympus recall portal. a. Go to https://olympusamerica.com/recall b. Enter the file (recall) number: 0430 4. If you have further distributed this product, identify your customers, forward them this notification, and appropriately document your notification process. If you require additional information, please contact Tara Safi, Field Corrective Actions Administrator at [email protected].
Quantity in commerce
522 units
Distribution
Worldwide distribution - US Nationwide and the countries of CN, DE, HK, IT, KR, SG, TH, TW.
Product codes and lots
Model-UDI: INS-5040 Always-On vTrack Universal Tracker, LG 7-12ga 00815686020705; INS-5039 Always-On vTrack Universal Tracker, SM/CL, 12-18ga 00815686020699; INS-5036 "22 UTW 125mm Always-On Tip Tracked Biopsy Needle, Trocar point" 00815686020361; INS-5034 17 UTW Lg 123mm Always-On Tip Tracked Biopsy Needle, Chiba point 00815686020354; INS-5032 "19 UTW 150mm Always-On Tip Tracked Biopsy Needle, Chiba point" 00815686020330; INS-5029 "19 UTW 105 mm Always-On Tip Tracked biopsy needle, Chiba point" 00815686020309; INS-5028 19 UTW 155 mm Always-On Tip Tracked biopsy needle, Chiba point 00815686020293; INS-5024 "17ga 150mm Always-On Tip Tracked Biopsy Needle, Chiba point" 00815686020255; INS-5023 17ga 100mm Always-On Tip Tracked Biopsy Needle, Chiba point 00815686020248; INS-5017 17ga x 155mm Tip Tracked Bx Needle 00815686020224; INS-5016 17ga x 155mm Tip Tracked Bx Needle 00815686020217; All serial numbers and all lot numbers

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2663-2023
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2659-2023Class IIOngoingVoluntary: Firm initiated
Z-2660-2023Class IIOngoingVoluntary: Firm initiated
Z-2661-2023Class IIOngoingVoluntary: Firm initiated
Z-2662-2023Class IIOngoingVoluntary: Firm initiated
Z-2663-2023Class IIOngoingVoluntary: Firm initiated
Z-2664-2023Class IIOngoingVoluntary: Firm initiated