TyTek Tension Pneumothorax Access Kit TPAK10: 10 gauge, 3.25 inch- Intended for introduction into the body to...
Tytek Medical Inc began a recall of TyTek Tension Pneumothorax Access Kit TPAK10: 10 gauge, 3.25 inch- Intended for introduction into the body to facilitate removal of air from the pleural cavity... on . FDA classified it as Class II and gives the reason as "Updated IFU to inlclude: (1) identify the risks associated with the hazard of needle decompression and (2) improve the artwork showing placement site location". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-2611-2023
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Tytek Medical Inc
- Product
- TyTek Tension Pneumothorax Access Kit TPAK10: 10 gauge, 3.25 inch- Intended for introduction into the body to facilitate removal of air from the pleural cavity in patients with a tension pneumothorax Model Number: TM-310
- Product code
- GAA Needle, Aspiration And Injection, Disposable
- Reason for recall
- Updated IFU to inlclude: (1) identify the risks associated with the hazard of needle decompression and (2) improve the artwork showing placement site location
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- TyTek New Product IFU was initially sent out to all who purchased the product on 08Dec2020 with notification on what was updated. An official field safety notice (FSN) was sent to customers on 02Aug2021.The changes to the IFUs were proactive in nature. There has been no change to the Products themselves. When used in a manner consistent with the IFUs, the Products are safe for their intended use, and we wish to reiterate that there have been no incidents whatsoever. For you reference I have attached the IFU that we originally sent you in December 2020. Please do not hesitate to contact us should you have any additional questions. The change in our IFUs for these products was implemented to: (1) identify the risks associated with the hazard of needle decompression and (2) improve the artwork showing placement site location.
- Quantity in commerce
- 4400 units
- Distribution
- OH, SC, VA Foreign: Country Australia Denmark Finland Germany Hong Kong/China Israel SPAIN The Netherlands Turkey UAE UK
- Product codes and lots
- UDI: 00855204008020 All lots produced prior to Oct 5, 2020
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2610-2023 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2611-2023 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2612-2023 | Class II | Terminated | Voluntary: Firm initiated |