Z-0007-2024Class I
3ML Syringe Luer Lock with needle 20GX1, Part Number 15-03ML-0, 100 per box, 24 boxes per case
Fresenius Medical Care Renal Therapies Group, LLC / initiated 2023-08-08 / Open, Classified / root cause: design
Fresenius Medical Care Holdings, Inc. began a recall of 3ML Syringe Luer Lock with needle 20GX1, Part Number 15-03ML-0, 100 per box, 24 boxes per case on . FDA classified it as Class I and gives the reason as "There is a potential for blood backfill leakage and foreign material". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0007-2024
- Classification
- Class I
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Fresenius Medical Care Renal Therapies Group, LLC
- Product
- 3ML Syringe Luer Lock with needle 20GX1, Part Number 15-03ML-0, 100 per box, 24 boxes per case
- Product code
- FMF Syringe, Piston
- Reason for recall
- There is a potential for blood backfill leakage and foreign material.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- An URGENT PRODUCT RECALL notification letter dated 8/8/23 was sent to customers. For return of unused product: Please check your stock. If you have any of the product(s) listed above, place the product in a secure, segregated area. Please contact FMCRTG Customer Service at 1-800-323-5188 for instructions on how to return the product and receive credit. o Reference to FA-2023-17-C RECALL o Specify if urgent replacement of product is needed when calling. NOTE: If you have no other product other than Sanxin, you may use in order to continue treatments until you receive replacement product. However, please monitor and be attentive of these potential issues prior to and immediately during use. Please sign and return the completed reply form (enclosed) for confirmation that you have received this notice. We apologize for any inconvenience this action may have caused and appreciate your cooperation.
- Quantity in commerce
- 124,773 boxes total
- Distribution
- US Nationwide.
- Product codes and lots
- UDI-DI: 16945060500101; All Lots
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0005-2024 | Class I | Ongoing | Voluntary: Firm initiated | ||
| Z-0006-2024 | Class I | Ongoing | Voluntary: Firm initiated | ||
| Z-0007-2024 | Class I | Ongoing | Voluntary: Firm initiated |