Z-2625-2023Class II

Philips Tempus Pro Patient Monitor, REF: 00-1004-R, 00-1007-R, 00-1024-R, 00-1026-R, 00-1004, and 00-1007...

Remote Diagnostic Technologies Ltd. / initiated 2023-08-24 / Open, Classified / root cause: design

Remote Diagnostic Technologies Ltd. began a recall of Philips Tempus Pro Patient Monitor, REF: 00-1004-R, 00-1007-R, 00-1024-R, 00-1026-R, 00-1004, and 00-1007, when Used With Tempus Pro Video Laryngoscope, REF... on . FDA classified it as Class II and gives the reason as "Full screen error message may occur on patient monitors with affected hardware version, either during or after laryngoscopes are unplugged from monitors, informing users a 60-100 second shutdown/restart are required...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.

Recall record

Recall number
Z-2625-2023
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Remote Diagnostic Technologies Ltd.
Product
Philips Tempus Pro Patient Monitor, REF: 00-1004-R, 00-1007-R, 00-1024-R, 00-1026-R, 00-1004, and 00-1007, when Used With Tempus Pro Video Laryngoscope, REF 01-2044
Product code
MHX Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)
Reason for recall
Full screen error message may occur on patient monitors with affected hardware version, either during or after laryngoscopes are unplugged from monitors, informing users a 60-100 second shutdown/restart are required. Error prevents: viewing graphical patient vital signs, sounding of pulse audio tone, visualization of patient airway used during intubation; may lead to treatment delay or hypoxia.
Root cause tag
design
FDA root cause
Device Design
Action
On 8/24/23 Correction Notices were mailed to customers and distributors who were asked to do the following: 1) If monitor has Trizeps-7 Hardware, remove the laryngoscope from service with this monitor. Users must use an alternative laryngoscope not connected to the Tempus Pro Monitor to manage the patient s airway to avoid interruption in patient care. The Tempus Pro Monitor can remain in service if the Tempus Pro Video Laryngoscope is not connected to the monitor. 2) If monitor has Trizeps-6 Hardware, the laryngoscope can continue to be used with the monitor. 3) Post this Urgent Medical Device Correction letter on or near the affected device. 4) Pass this notice on to all those who need to be aware within your organization or to any organization where the potentially affected devices have been transferred (if appropriate). 5) Complete and return the Urgent Medical Device Correction response form. Distributors were asked to take the following additional actions: If you have any Tempus Pro Video Laryngoscopes or any Tempus Pro Monitors in stock that have not been distributed, do not distribute them. Modify the response form to include your firm's contact information and then send the response form along with the customer correction notice to customers. If you need any further information or support contact the firm at (800) 722-9377 or email [email protected]. On 11/6/23, updated correction notices were distributed adding two additional models that could become affected if their hardware is updated. Additionally, the notices asked customers to install software (V07.34) on Tempus Pro Monitors according to installation instructions.
Quantity in commerce
3,524
Distribution
US:AZ, AL, CA, CO, DC, DE, FL , GA, HI, IA, ID, IL, IN , KS, KY, MA , MD , ME, MI, MN, MO, MS, MT, NC , ND, NE, NJ, NM , NV, NY, OH, OK, OR, PA, TN, TX, UT, VA , WA, WI, WV, WY. OUS: AE, AT, AU, BE, BN, BO, CA, CH, CZ, DE, DK, ES, FR, GB, HK, IE, IL, IS, IT, KY, LU, MY, NL, NO, OM, PT, RO, SA, SE, SG, SI, TH, YT, ZA
Product codes and lots
Monitors with hardware version: Trizeps-7, and software versions v07.26 through v07.30. REF/UDI-DI: 00-1004-R/05060472440020, 05060472442901; 00-1007-R/05060472440013, 05060472442918; 00-1024-R/05060472441027, 05060472442925; 00-1026-R/05060472441058, 05060472442932; 00-1004/05060472440624; 00-1007/05060472440655

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2625-2023
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2625-2023Class IIOngoingVoluntary: Firm initiated