Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2653-2023Class II

Apellis Injection Kit 29g Injection Needle -single use intended to inject fluids into, or withdraw fluids...

Apellis / initiated 2023-08-22 / Open, Classified

Apellis Pharmaceuticals, Inc. began a recall of Apellis Injection Kit 29g Injection Needle -single use intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin... on . FDA classified it as Class II and gives the reason as "19-gauge x 1" inch filter needles showed appearances of a small fraction of non-preferred characteristics. A causal relationship has not been established between the structural variations in this 19-gauge filter needle...". As of , the recall status on record is Open, Classified.

Recall record

Recall number
Z-2653-2023
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Apellis
Product
Apellis Injection Kit 29g Injection Needle -single use intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin and either an 18-gauge or a 19-gauge 5 micron filter needle. Ref: IVT-KIT-29G
Product code
FMI Needle, Hypodermic, Single Lumen
Reason for recall
19-gauge x 1" inch filter needles showed appearances of a small fraction of non-preferred characteristics. A causal relationship has not been established between the structural variations in this 19-gauge filter needle and the rare events of retinal vasculitis
FDA root cause
Under Investigation by firm
Action
Apellis issued Medical Device Field Correction letters via email on 8/22/23 to Healthcare Practitioners and Distributors/Specialty Distributors and Pharmacy/ Logistics. The communication to the Healthcare Practitioners also includes update on the rare events of retinal vasculitis with real world use of SYFOVRE (pegcetacoplan injection). A causal relationship has not been established between the 19-gauge needle and the rare events of retinal vasculitis in the real world. Letter states reason for recall, health risk and action to take: Actions to be taken by the Customer/Distributor: Please note that as part of this field action, you are requested to: " Confirm your availability and participation by sending back an email to acknowledge receipt of this message. " Immediately cease distribution of the below injection kit lots and place any remaining inventory of the lots On Hold. " Provide a summary of current on-hand inventory of the injection kit lots below. " Provide a distribution report of the injection kit lots including customer name/address, lot shipped, quantity shipped, and date. Please provide the Distribution report in a maximum time of four business hours. Apellis will also be recommending that practitioners discontinue use of, and dispose, any remaining injection kits that contain the specific 19-gauge filter needle and recommend they instead use the injection kits containing the 18-gauge filter needle already in distribution. Apellis is now exclusively distributing injection kits with the 18-gauge filter needle. If you need to place a supplemental order to replace quarantined lots, please place them immediately through your normal ordering process and Apellis will expedite shipping replacement injection kits./ Apellis will also be notifying physicians that if they do not have an injection kit with the 18-gauge filter needle, they may request a new kit at no cost by emailing [email protected] or contacting their Apellis Territory Business Mana
Quantity in commerce
61,943 units
Distribution
US Nationwide distribution.
Product codes and lots
UDI-DI: 00860008043672 Kit Lots: 223186 230036 230056 230236 230316 230326 230336 230406 231926 231946 231956 231986

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2653-2023
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2653-2023Class IIOngoingVoluntary: Firm initiated