Z-0075-2024Class I

Olympus High-Flow Insufflation Unit, Model UHI-4. For laparoscopic surgery

OLYMPUS Corporation of the Americas / initiated 2023-09-22 / Open, Classified

Olympus Corporation of the Americas began a recall of Olympus High-Flow Insufflation Unit, Model UHI-4. For laparoscopic surgery on . FDA classified it as Class I and gives the reason as "There have been reports of patients suffering arrhythmias, reported as short cardiac arrests, during surgical procedures where UHI-4s were used. These events may have been due to an over insufflation of the abdominal...". As of , the recall status on record is Open, Classified.

Recall record

Recall number
Z-0075-2024
Classification
Class I
Status
Open, Classified
Initiated
Posted
Recalling firm
OLYMPUS Corporation of the Americas
Product
Olympus High-Flow Insufflation Unit, Model UHI-4. For laparoscopic surgery.
Product code
HIF Insufflator, Laparoscopic
Reason for recall
There have been reports of patients suffering arrhythmias, reported as short cardiac arrests, during surgical procedures where UHI-4s were used. These events may have been due to an over insufflation of the abdominal cavity resulting from use of the UHI-4 during the procedures.
FDA root cause
Under Investigation by firm
Action
An URGENT: MEDICAL DEVICE CORRECTIVE ACTION notification letter dated 9/22/23 was sent to customers. Action steps to be taken by the end user: Our records indicate that your facility has purchased one or more of the Olympus UHI-4. Olympus requires you to take the following actions: 1. Carefully read the content of this Field Corrective Action Letter. 2. Ensure all personnel are completely knowledgeable and thoroughly trained on the content of this letter. 3. Complete the enclosed response form and return it to our recall partner, Sedgwick, via e-mail at [email protected] or by fax at 866-367-6717. For any questions about the acknowledgement form, please call the Sedgwick team at 855-215-4972. 4. If you have further distributed this product, identify your customers, and forward them this Letter. Please appropriately document your notification process and let us know the end-customer feedback accordingly. Olympus requests that you report complaints, including any injuries associated with over insufflation during the procedure with UHI-4, to Olympus. Please report complaints to Technical Assistance Center (TAC) at 1-800-848-9024, option 1. Adverse events experienced with the use of this product may also be reported to the Food and Drug Administration MedWatch Adverse Event Reporting program either online, by regular mail or by fax. An URGENT: MEDICAL DEVICE CORRECTIVE ACTION - UPDATE notification letter dated 10/25/23 was sent to customers. Action steps to be taken by the end user: Our records indicate that your facility has purchased one or more of the Olympus UHI-4. Olympus requires you to take the following actions: 1. Inspect your inventory and identify any UHI-4 High flow insufflation unit(s). Quarantine and mark your units appropriately by your site to prevent usage. If your facility does not have alternatives or is unable to obtain alternatives, you may choose to use the UHI-4 while using extreme caution and after reviewing the information above
Quantity in commerce
3136 units
Distribution
US Nationwide distribution.
Product codes and lots
UDI-DI: 04953170324147; All Serial Numbers

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0075-2024
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0075-2024Class IOngoingVoluntary: Firm initiated