Z-0124-2024Class I

8713052U, Infusomat Large Volume Pump, Non-Wireless BATTERY PACK; Volumetric Infusion Pump System

B Braun Medical Inc / initiated 2023-09-21 / Open, Classified / root cause: design

B. Braun Medical, Inc. began a recall of 8713052U, Infusomat Large Volume Pump, Non-Wireless BATTERY PACK; Volumetric Infusion Pump System on . FDA classified it as Class I and gives the reason as "Upstream occlusion alarm may sound when no occlusion exists, and the device will stop pumping. Interruption to the infusion of high-risk medications may lead to hemodynamic instability requiring medical intervention to...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.

Recall record

Recall number
Z-0124-2024
Classification
Class I
Status
Open, Classified
Initiated
Posted
Recalling firm
B Braun Medical Inc
Product
8713052U, Infusomat Large Volume Pump, Non-Wireless BATTERY PACK; Volumetric Infusion Pump System
Product code
FRN Pump, Infusion
Reason for recall
Upstream occlusion alarm may sound when no occlusion exists, and the device will stop pumping. Interruption to the infusion of high-risk medications may lead to hemodynamic instability requiring medical intervention to prevent permanent impairment to body structures or body functions. In some cases, this may be life-threatening or may lead to death.
Root cause tag
design
FDA root cause
Component design/selection
Action
On September 21, 2023, the firm began notifying customers of the issue via "URGENT MEDICAL DEVICE CORRECTION" letters. BBMI will schedule service to update impacted devices. Customers were provided with stickers to identify affected units while waiting for the service to be completed. In the meantime, customers were informed of measures that may aid in continued use of the pumps, such as only using non-impacted devices for the delivery of high-risk medications and placing the impacted units in care areas that do not administer high risk medications. Customers may contact BBMI by calling 1-800-627-7867 (1-800-627-PUMP) for technical support for potential alternate interim measures. ***Updated 8/7/24*** On August 7, 2024, firm identified additional units which were serviced using potentially impacted sensors. Serial numbers were added to the two models already in scope (Article Numbers 8713051U and 8713052U). Serial numbers from an additional model (Article No. 8713050U) also added and tracked under RES 95093.
Quantity in commerce
884
Distribution
Domestic distribution nationwide. Foreign distribution to Canada.
Product codes and lots
UDI-DI: 04046964708626 Serial Numbers: 878498 -881897 ***Update 8/7/24*** Additional SNs 133238, 133975, 490442, 546815, 559325, 561645, 562713, 593049, 617405, 617409, 618522, 783680, 810031