Z-0205-2024Class II

Smiths Medical Portex Bivona Pediatric Tracheostomy Tube, Cuffless, Model 60P045, Sterile

Abbott Medical / initiated 2023-08-22 / Open, Classified / root cause: labeling and UDI

Smiths Medical ASD Inc. began a recall of Smiths Medical Portex Bivona Pediatric Tracheostomy Tube, Cuffless, Model 60P045, Sterile on . FDA classified it as Class II and gives the reason as "The unit box is labeled with an incorrect expiration date (the manufacturing date of 3/17/2021) but the individual packaging contains the correct expiration date (3/17/2026)". As of , the recall status on record is Open, Classified, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-0205-2024
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Abbott Medical
Product
Smiths Medical Portex Bivona Pediatric Tracheostomy Tube, Cuffless, Model 60P045, Sterile.
Product code
JOH Tube Tracheostomy And Tube Cuff
Reason for recall
The unit box is labeled with an incorrect expiration date (the manufacturing date of 3/17/2021) but the individual packaging contains the correct expiration date (3/17/2026).
Root cause tag
labeling and UDI
FDA root cause
Labeling mix-ups
Action
On 8/22/2023, a third-party recall notification service issued recall notifications to the direct accounts on behalf of the recalling firm via UPS ground. The letter explained the reason for the notification and embedded pictures of the incorrect and correct labels into the letter. The letter also provided the potential risk and required actions for users. The potential risk involves the potential for a customer to assess the product as expired, resulting in the potential for a delay in the availability to use the tracheostomy device. If the customer did not identify the incorrect device unit carton/secondary packaging labeling upon receipt and provided the items to clinicians, the individual device tray/primary packaging would still allow correct identification of the expiration date at the point of use, as the expiry date on the individual unit is correctly labeled. The required actions included (1) informing potential users within their organization of the notification and to complete and return the attached response form via email within ten days of receipt to acknowledge if they have the affected product and/or if it has already been used; and (2) if the account has further distributed the product, immediately notify those customers of the field correction. The letter instructs that if the product has not already been used, the product must be used before the correct expiry of March 17, 2026.
Distribution
Distribution was made to AZ, CA, FL, HI, IL, IN, MA, MN, MO, MS, NC, NJ, OH, PA, SC, TX, VA, and WA. There was no government/military distribution. Foreign distribution was made to Canada.
Product codes and lots
Lot #4125816, exp. 3/17/2026, UDI-DI 15021312005769.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0205-2024
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0205-2024Class IIOngoingVoluntary: Firm initiated