Z-2520-2023Class II

Incisive CT, software version 5.0

Philips North America LLC / initiated 2023-08-02 / Open, Classified / root cause: software

Philips Healthcare began a recall of Incisive CT, software version 5.0 on . FDA classified it as Class II and gives the reason as "Philips has identified three software issues with compliance concerns to Subchapter J: Issue 1: Coronary Computed Tomography Angiography (CTA) scan may not be triggered at the optimal time. Issue 2: Use of aborted...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.

Recall record

Recall number
Z-2520-2023
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Philips North America LLC
Product
Incisive CT, software version 5.0
Product code
JAK System, X-Ray, Tomography, Computed
Reason for recall
Philips has identified three software issues with compliance concerns to Subchapter J: Issue 1: Coronary Computed Tomography Angiography (CTA) scan may not be triggered at the optimal time. Issue 2: Use of aborted surview for scan planning will result in deviation from intended scan area. Issue 3: Actual clinical scan with bariatric (Noah) couch will deviate from the intended scan area if surview scan length is greater than 1832mm.
Root cause tag
software
FDA root cause
Radiation Control for Health and Safety Act
Action
Philips will inform customers of the issue via an Electronic Product Radiation Defect notification. Philips will schedule an appointment with customers to install the software update.
Quantity in commerce
140 systems
Distribution
US Nationwide Distribution
Product codes and lots
software version 5.0

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2520-2023
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2520-2023Class IIOngoingFDA Mandated