LFS Flat screen arm system with surgical lamp for Models Aurora series, Aurora 2 series, Aurora 3 series...
DKK Dai-Ichi Shomei Co., Ltd. began a recall of LFS Flat screen arm system with surgical lamp for Models Aurora series, Aurora 2 series, Aurora 3 series, Aurora 4 series, Aurora Astro series, Stellar XL, and... on . FDA classified it as Class II and gives the reason as "The stopper block on the LFS bracket could be broken or missing. Without this stopper block, the flat screen attached to the LFS flat screen system can rotate beyond the original angle limit and it could cause excessive...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0071-2024
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- DKK Dai-Ichi Shomei Co., Ltd.
- Product
- LFS Flat screen arm system with surgical lamp for Models Aurora series, Aurora 2 series, Aurora 3 series, Aurora 4 series, Aurora Astro series, Stellar XL, and Stellar series, LFS Flat Screen Arm System Model Numbers LFSST2323, LFSLFSST29TV23, LFSST29TV23, LFSST23TV, LFSST23, LFSST2323TV, LFSLFSST2323, LDFSLFSLED7TV5, LDFSLFSLED7TV55, LFSLED55, LFSLFSST23, LFSST2923TV, LFSST29TV, LFSST29, LFSST19WC, LFSST2923, LFSLFSST-1, LFSST292323, LFSST1919WC, LFS ST2319WC
- Product code
- FSY Light, Surgical, Ceiling Mounted
- Reason for recall
- The stopper block on the LFS bracket could be broken or missing. Without this stopper block, the flat screen attached to the LFS flat screen system can rotate beyond the original angle limit and it could cause excessive force to be given to the spring shaft of the LFS bracket. Due to the excessive force given to the spring shaft, the spring shaft could break and shear off.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- An URGENT RECALL notification letter dated 8/23/23 was sent to the customer. Actions to be Taken by the Manufacturer We will provide an additional metal cover for the LFS bracket so that even if the shaft shears off, it will not pop out from the cover as well as instructions to properly install the cover. Also, the additional metal cover provides additional support to the LFS bracket and prevents the stopper block from broken or missing even if the excessive force is added to the bracket. The physical correction will be announced later. Other Information Contact information for questions: TEL: +81-3-3969-9221 FAX: +81-3-3969-9269 Email: [email protected] Monday through Friday, 8:00 AM to 5:00 PM, Japan Standard Time.
- Quantity in commerce
- 256 units
- Distribution
- US States: MI
- Product codes and lots
- Serial Numbers: 06087323 - 06087352, 06087353 - 06087354, 06081292 - 06081293, 06087448 - 06087449, 06087450 - 06087469, 06097595 - 06097596, 06107716 - 06107718, 06107719 - 06107728, 06107730, 06107731 - 06107732, 06111348 - 06111349, 06117855 - 06117856, 06117857 - 06117866, 06117868 - 06117868, 06111350 - 06111351, 06117949 - 06117949, 06117954 - 06117973, 06118031 - 06118032, 06128116 - 06128131, 06128172 - 06128176, 06128177 - 06128186, 06128187 - 06128188, 07010004, 07010002, 07010005 - 07010006, 07028436 - 07028441, 07030080 - 07030081, 07048491 - 07048495, 07058506 - 07058507, 07058508 - 07058512, 07068602 - 07068604, 07068605 - 07068607, 07068611 - 07068622, 07068609 - 07068610, 07068667 - 07068670, 07061479 - 07061484, 07078679 - 07078680, 07071485 - 07071488, 07071501, 07088741 - 07088742, 07081124, 07088766, 07098769 - 07098773, 07098767 - 07098768, 07098774 - 07098775, 07108885 - 07108889, 07108883 - 07108884, 07108876 - 07108882, 07108890 - 07108891, 07101533 - 07101534, 07129027 - 07129028, RA000182, RA000184, 08029174 - 08029183, 08029187 - 08029188, 08029189 - 08029190, 09111694, 10011776, 10011777, 10011778, 10011779, 10011780, 10011781, 10011782, 10051792, 10051794, 10051796, 10051798, 10051800, 10051802, 10051804, 100610044, 100610045, 100610046, 100610047, 100610048, 100610049, 100610050, 100610051, 100610052, 100610053, 101010180, 101010182, 101010184, 101010186, 101010188, 101010190, 101010192, 101210252, BM005435 - BM005435, BM006121 - BM006121, BM008121 - BM008127
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0071-2024 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0072-2024 | Class II | Ongoing | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.