Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0126-2024Class II

SynCardia 50cc TAH-t, Clinical, REF 570500, Sterile EO

SynCardia Systems LLC / initiated 2023-08-01 / Completed

SynCardia Systems LLC began a recall of SynCardia 50cc TAH-t, Clinical, REF 570500, Sterile EO on . FDA classified it as Class II and gives the reason as "Artificial hearts contain a epoxy resin that has not been reviewed by the FDA for safety and efficacy". As of , the recall status on record is Completed.

Recall record

Recall number
Z-0126-2024
Classification
Class II
Status
Completed
Initiated
Posted
Recalling firm
SynCardia Systems LLC
Product
SynCardia 50cc TAH-t, Clinical, REF 570500, Sterile EO
Product code
LOZ Artificial Heart
Reason for recall
Artificial hearts contain a epoxy resin that has not been reviewed by the FDA for safety and efficacy.
FDA root cause
Other
Action
On August 22, 2023, SynCardia issued an Urgent Medical Device Correction Notification to all affected consignees via Email. SynCardia asked consignees to take the following actions: 1. You may continue to use the device while the change undergoes FDA review for safety and effectiveness. 2.Ensure that all medical professionals in your institution are informed of this Urgent Medical Device Correction. 3. Please complete the Customer Acknowledgement Form and return it via email to SynCardia Systems, LLC, within five business days of receipt of this letter. 4.For patients with implanted devices affected by this correction, monitor patients closely and take appropriate actions you believe necessary if any issues arise. 5. Adverse reactions or quality problems experienced with this product may be reported to the FDA s MedWatch Adverse Event Reporting program either online, by regular mail, or by fax. " Complete and submit the report online: www.fda.gov/medwatch/report.htm " Regular Mail or Fax: Download the form from www.fda.gov/medwatch/getforms.htm or call 1-800-322-1088 to request a reporting form, then complete and return to the address on the pre-addressed form, or submit by fax to 1-800-FDA-0178
Quantity in commerce
114 units
Distribution
US: FL, WI, AZ, TX, CA, OH, NC, IN, OK, IL, VA, NY, NE, PA, WA, OUS: Saudi Arabia, Austria, Italy, Macedonia, Turkey, Germany, France, Kuwait, Serbia, United Kingdom, Canada, Lithuania, Slovenia Australia
Product codes and lots
All Lots/ UDI: 00858000003053

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0126-2024
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0126-2024Class IICompletedVoluntary: Firm initiated