Z-0126-2024Class II
SynCardia 50cc TAH-t, Clinical, REF 570500, Sterile EO
SynCardia Systems LLC / initiated 2023-08-01 / Completed
SynCardia Systems LLC began a recall of SynCardia 50cc TAH-t, Clinical, REF 570500, Sterile EO on . FDA classified it as Class II and gives the reason as "Artificial hearts contain a epoxy resin that has not been reviewed by the FDA for safety and efficacy". As of , the recall status on record is Completed.
Recall record
- Recall number
- Z-0126-2024
- Classification
- Class II
- Status
- Completed
- Initiated
- Posted
- Recalling firm
- SynCardia Systems LLC
- Product
- SynCardia 50cc TAH-t, Clinical, REF 570500, Sterile EO
- Product code
- LOZ Artificial Heart
- Reason for recall
- Artificial hearts contain a epoxy resin that has not been reviewed by the FDA for safety and efficacy.
- FDA root cause
- Other
- Action
- On August 22, 2023, SynCardia issued an Urgent Medical Device Correction Notification to all affected consignees via Email. SynCardia asked consignees to take the following actions: 1. You may continue to use the device while the change undergoes FDA review for safety and effectiveness. 2.Ensure that all medical professionals in your institution are informed of this Urgent Medical Device Correction. 3. Please complete the Customer Acknowledgement Form and return it via email to SynCardia Systems, LLC, within five business days of receipt of this letter. 4.For patients with implanted devices affected by this correction, monitor patients closely and take appropriate actions you believe necessary if any issues arise. 5. Adverse reactions or quality problems experienced with this product may be reported to the FDA s MedWatch Adverse Event Reporting program either online, by regular mail, or by fax. " Complete and submit the report online: www.fda.gov/medwatch/report.htm " Regular Mail or Fax: Download the form from www.fda.gov/medwatch/getforms.htm or call 1-800-322-1088 to request a reporting form, then complete and return to the address on the pre-addressed form, or submit by fax to 1-800-FDA-0178
- Quantity in commerce
- 114 units
- Distribution
- US: FL, WI, AZ, TX, CA, OH, NC, IN, OK, IL, VA, NY, NE, PA, WA, OUS: Saudi Arabia, Austria, Italy, Macedonia, Turkey, Germany, France, Kuwait, Serbia, United Kingdom, Canada, Lithuania, Slovenia Australia
- Product codes and lots
- All Lots/ UDI: 00858000003053
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0126-2024 | Class II | Completed | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.