Constellation Vision System, REF: 8065751150
Alcon Research, LLC began a recall of Constellation Vision System, REF: 8065751150 on . FDA classified it as Class II and gives the reason as "Ophthalmic microsurgical vision systems have a new Printed Circuit Board (PCB) design that could lead to higher voltage to a PCB component, that could cause component failure and loss of pressure control for the use of...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0099-2024
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Alcon Laboratories
- Product
- Constellation Vision System, REF: 8065751150
- Product code
- HQC Unit, Phacofragmentation
- Reason for recall
- Ophthalmic microsurgical vision systems have a new Printed Circuit Board (PCB) design that could lead to higher voltage to a PCB component, that could cause component failure and loss of pressure control for the use of cutters, scissors, and forceps, which could lead to an increased risk of microbial infection, tissue damage and progressive visual impairment.
- Root cause tag
- design
- FDA root cause
- Component change control
- Action
- Starting on 8/18/23, Field Service Engineers told customers their devices needed pneumatics module replacements, and this replacement activity was subsequently completed. On 9/29/23, correction notices were mailed to customers informing them vision systems were corrected and asking they complete and return the response form to [email protected] Should you have any questions or concerns call Customer Service at 1-800-862-5266.
- Quantity in commerce
- 3
- Distribution
- US: GA, IL, PR
- Product codes and lots
- UDI-DI: 00380657511501, Serial Numbers: 1403028701X, 1202865201X, 0802983301X
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0099-2024 | Class II | Ongoing | Voluntary: Firm initiated |