Z-0093-2024Class II

BD Pyxis CII Safe ES, REF: 1116-00; and BD Pyxis : CII Safe ES Desktop PC, Wired SCNR, REF: 107-254-01; CII...

CareFusion 303, Inc / initiated 2023-09-06 / Open, Classified / root cause: software

CareFusion 303, Inc. began a recall of BD Pyxis CII Safe ES, REF: 1116-00; and BD Pyxis : CII Safe ES Desktop PC, Wired SCNR, REF: 107-254-01; CII Safe ES Desktop PC, W-LESS SCNR, BIO, REF... on . FDA classified it as Class II and gives the reason as "When global edit is used to update multiple formulary properties simultaneously, the following properties: 1) Require Lot Number on Recall, 2) Access Destruction Bin (Witness), 3) Add to Destruction Bin (Count/Empty)...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.

Recall record

Recall number
Z-0093-2024
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
CareFusion 303, Inc
Product
BD Pyxis CII Safe ES, REF: 1116-00; and BD Pyxis : CII Safe ES Desktop PC, Wired SCNR, REF: 107-254-01; CII Safe ES Desktop PC, W-LESS SCNR, BIO, REF: 107-255-01; Enterprise Server, REF: 1115-00, 134157-01, 137670-02, 134148-01, 135292-01, 136449-01, 133760-01
Product code
BRY Cabinet, Table And Tray, Anesthesia
Reason for recall
When global edit is used to update multiple formulary properties simultaneously, the following properties: 1) Require Lot Number on Recall, 2) Access Destruction Bin (Witness), 3) Add to Destruction Bin (Count/Empty), 4) CII Safe Stock Out Notice are, without user warning, overwritten to the value NO , which could result in diversion of controlled substances from the automated dispensing cabinet.
Root cause tag
software
FDA root cause
Software change control
Action
On 9/6/23, correction notices were mailed to customers informing them to do the following: 1) Do not use the global edit feature until your software is updated. 2) If a global edit has already been performed, it is recommended to verify that the formulary device settings are accurate for the 4 configuration settings. Controlled substances that have been edited via the global edit feature should be reviewed retrospectively for discrepancies in the destruction bin and a manual witness record may be needed until the settings are adjusted for both "Access Destruction Bin" and "Add to Destruction Bin (count/empty)." 3) Disseminate this notification within your facility network to ensure they are aware of this issue. 4) Complete the attached Customer Response Form and return via email to [email protected] Software patch for v.1.7.3 was released on July 20, 2023, and commercial release software v.1.7.4 was released for on July 31, 2023. Firm will contact customers after the receipt of completed response forms to schedule times to install software update. The following contacts were provided: Technical Support, 1-800-727-6102, 24 hours, 7 days a week; Customer Self Service Portal, https://service.carefusion.com; North American Regional Complaint Center, 1-844-823-5433, Say "Recall" when prompted, Mon-Fri 8:00am and 5:00pm CT, email: [email protected]
Quantity in commerce
61
Distribution
US Nationwide distribution in the states of LA, NC, NY, MI, MA, NV, KS, IA, CA.
Product codes and lots
REF: 1116-00, UDI-DI: 10885403512605. Software Versions: 1.7.3 and 1.7.4 (Limited commercial release)

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0093-2024
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0093-2024Class IIOngoingVoluntary: Firm initiated