Multichem P, Part Code 08P90-10
TECHNO-PATH MANUFACTURING LTD. began a recall of Multichem P, Part Code 08P90-10 on . FDA classified it as Class II and gives the reason as "Technopath Manufacturing Ltd. has identified an issue with the glass vial commodity used to fill the Multichem S Plus Level 1, Multichem S Plus Level 2, Multichem S Plus Level 3, Multichem IA Plus and Multichem P...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0106-2024
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Techno-Path Manufacturing Ltd.
- Product
- Multichem P, Part Code 08P90-10
- Product code
- JJY Multi-Analyte Controls, All Kinds (Assayed)
- Reason for recall
- Technopath Manufacturing Ltd. has identified an issue with the glass vial commodity used to fill the Multichem S Plus Level 1, Multichem S Plus Level 2, Multichem S Plus Level 3, Multichem IA Plus and Multichem P Controls. Some customers have experienced broken vials when receiving or thawing Technopath Multichem IA Plus and Multichem S Plus Controls. Breakages have been observed upon delivery, while thawing or while opening the vial. The breakages can result in the bottom of the vial breaking off, breakage in the neck thread area or cracking on the sides of the vial. This may expose the user to biohazardous material if cut and may cause a delay in patient results.
- Root cause tag
- manufacturing process
- FDA root cause
- Unknown/Undetermined by firm
- Action
- LGC Group notified their sole US consignee (distribution partner) of the recall on 08/25/2023 via email. The notice explained the issue and the risk with the request to notify their end users. The end user was advised to take additional care when handling the glass vials due to the potential of breakages upon receipt, while thawing, or while opening the vials.
- Quantity in commerce
- 2964 units
- Distribution
- US Nationwide distribution in the state of Illinois.
- Product codes and lots
- UDI/DI 05391523440769, Lot Numbers: 04530621P, 04561021P, 04610523P
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0104-2024 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0105-2024 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0106-2024 | Class II | Ongoing | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.