Z-0112-2024Class II
Monoject(TM) 1 mL Tuberculin Syringe Luer-Lock Tip Soft Pack, Model Number/
Medtronic, Inc. / initiated 2023-09-13 / Open, Classified
Cardinal Health 200, LLC began a recall of Monoject(TM) 1 mL Tuberculin Syringe Luer-Lock Tip Soft Pack, Model Number/Product Code: 1180100777 on . FDA classified it as Class II and gives the reason as "Syringe contains conical tip that is not compatible with certain needleless IV connector(s)". As of , the recall status on record is Open, Classified.
Recall record
- Recall number
- Z-0112-2024
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Medtronic, Inc.
- Product
- Monoject(TM) 1 mL Tuberculin Syringe Luer-Lock Tip Soft Pack, Model Number/Product Code: 1180100777
- Product code
- FMF Syringe, Piston
- Reason for recall
- Syringe contains conical tip that is not compatible with certain needleless IV connector(s).
- FDA root cause
- Process change control
- Action
- Cardinal Health notified consignees on about 09/13/2023 via letter titled "URGENT MEDICAL DEVICE PRODUCT CORRECTION." Customers were instructed to review inventory for the affected product code and lot, ensure all personnel are informed cease utilizing the affected product with needleless IV connectors, post a copy of the notification in the storeroom where the product is stored, and notify any customers if the product has been further distributed/forwarded. Additionally, customers were asked to complete and return the included acknowledgment form.
- Quantity in commerce
- 3.927,994 units
- Distribution
- Nationwide distribution to AK, AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, WI, WV, WY
- Product codes and lots
- Model Number/Product Code: 1180100777. UDI/DI: 10192253034530 each, 20192253034537 box, 50192253034538 case. Lot Numbers: 221201, 221202, 221203, 230201, 230202, 230203, 230204, 230205, 230601.