Z-0142-2024Class II

Centricity Universal Viewer Workflow Manager, Image processing radiological system

GE Healthcare / initiated 2023-09-08 / Open, Classified / root cause: software

GE Healthcare began a recall of Centricity Universal Viewer Workflow Manager, Image processing radiological system on . FDA classified it as Class II and gives the reason as "When using Universal Viewer or Centricity Universal Viewer with Workflow Manager (WFM), if a third-party reporting application is used to launch exams, it could display information on a different patient than WFM and...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.

Recall record

Recall number
Z-0142-2024
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
GE Healthcare
Product
Centricity Universal Viewer Workflow Manager, Image processing radiological system
Product code
LLZ System, Image Processing, Radiological
Reason for recall
When using Universal Viewer or Centricity Universal Viewer with Workflow Manager (WFM), if a third-party reporting application is used to launch exams, it could display information on a different patient than WFM and the Viewer.
Root cause tag
software
FDA root cause
Software design
Action
GE Healthcare notified consignees on about 09/08/2023 via letter. Consignees were instructed that until the software update is available the device can still be utilized through an updated browser setting or through discontinuing the use of bi-directional reporting. A GE Healthcare representative will contact each consignee to arrange for the correction. Consignees were instructed to complete and return the provided acknowledgement response.
Quantity in commerce
85 units
Distribution
Nationwide distribution to AR, AZ, CA, CO, DE, FL, GA, IL, IN, ME, MI, MN, MO, NE, NJ, NM, NV, NY, OH, OK, PA, SC, SD, TN, TX. International distribution to Austria, Canada, France, Germany, Hong Kong, Israel, Italy, Philippines, Singapore, Thailand, Turkey, Vietnam.
Product codes and lots
UDI/DI 00840682145794, Software versions 7.0 SP1.1, 7.0 SP1, 7.0 SP0.0.5, 7.0 SP0.1.0. All product IDs with an affected software version are impacted.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0142-2024
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0142-2024Class IIOngoingVoluntary: Firm initiated
Z-0143-2024Class IIOngoingVoluntary: Firm initiated