Universal Viewer Workflow Manager, Image processing radiological system
GE Healthcare began a recall of Universal Viewer Workflow Manager, Image processing radiological system on . FDA classified it as Class II and gives the reason as "When using Universal Viewer or Centricity Universal Viewer with Workflow Manager (WFM), if a third-party reporting application is used to launch exams, it could display information on a different patient than WFM and...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0143-2024
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- GE Healthcare
- Product
- Universal Viewer Workflow Manager, Image processing radiological system
- Product code
- LLZ System, Image Processing, Radiological
- Reason for recall
- When using Universal Viewer or Centricity Universal Viewer with Workflow Manager (WFM), if a third-party reporting application is used to launch exams, it could display information on a different patient than WFM and the Viewer.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- GE Healthcare notified consignees on about 09/08/2023 via letter. Consignees were instructed that until the software update is available the device can still be utilized through an updated browser setting or through discontinuing the use of bi-directional reporting. A GE Healthcare representative will contact each consignee to arrange for the correction. Consignees were instructed to complete and return the provided acknowledgement response.
- Quantity in commerce
- 57 units
- Distribution
- Nationwide distribution to AR, AZ, CA, CO, DE, FL, GA, IL, IN, ME, MI, MN, MO, NE, NJ, NM, NV, NY, OH, OK, PA, SC, SD, TN, TX. International distribution to Austria, Canada, France, Germany, Hong Kong, Israel, Italy, Philippines, Singapore, Thailand, Turkey, Vietnam.
- Product codes and lots
- UDI/DI 00195278379610, Software versions 7.0 SP1.1, 7.0 SP1, 7.0 SP0.0.5, 7.0 SP0.1.0. All product IDs with an affected software version are impacted.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0142-2024 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0143-2024 | Class II | Ongoing | Voluntary: Firm initiated |